Capillary zone electrophoresis: Effect of physical parameters on separation efficiency and quantitation
Authors:
- 1 August 1985
- journal article
- research article
- Published by Wiley in Journal of High Resolution Chromatography
- Vol. 8 (8) , 407-411
Abstract
Separation efficiency in capillary zone electrophoresis was examined as a function of capillary length and diameter as well as solute concentration for a borosilicate glass capillary/phosphate buffer system. Electroosmotic flow coefficients in borosilicate, fused silica, and teflon capillaries were measured over the useful operating pH range of 3–8 at constant applied power (0.333 W). Sample introduction by a simple, low‐dispersion electromigration method was found to be a reliable procedure for solute quantitation.Keywords