Abstract
We have developed laser-diode-based optical-resolution photoacoustic microscopy (LD-OR-PAM) of superficial microvasculature which has the desirable properties of being compact, low-cost, and label-free. A 300-mW visible pulsed laser diode was operated at a 405 ± 5 nm wavelength with a pulse energy as low as 52 nJ. By using a 3.6 MHz ultrasound transducer, the system was tested on carbon fibers with a lateral resolution of 0.95 µm and an SNR of 38 dB. The subcutaneous microvasculature on a mouse back was imaged without an exogenous contrast agent which demonstrates the potential of the proposed prototype for skin chromophores. Our eventual goal is to offer a practical and affordable multi-wavelength functional LD-OR-PAM instrument suitable for clinical applications.
Funding Information
  • National Institutes of Health (NIH) (P41-EB015890, R01EY-021529, R01HL-125084, R01HL-127271)
  • National Natural Science Foundation of China (NSFC) (11304129)
  • Natural Science Foundation of Jiangxi Province (20143BBM26053, 20151BAB217025)