Energy Harvesting and Sensing With Embedded Piezoelectric Ceramics in Knee Implants
- 15 January 2018
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE/ASME Transactions on Mechatronics
- Vol. 23 (2) , 864-874
Abstract
The knee replacement is one of the most common orthopedic surgical interventions in the United States; however, recent studies have shown up to 20% of patients are dissatisfied with the outcome. One of the key issues to improving these operations is a better understanding of the ligamentous balance during and after surgery. The goal of this paper is to investigate the feasibility of embedding piezoelectric transducers in the polyethylene bearing of a total knee replacement to act as self-powered sensors to aid in the alignment and balance of the knee replacement by providing intra- and postoperative feedback to the surgeon. A model consisting of a polyethylene disc with a single embedded piezoelectric ceramic transducer is investigated as a basis for future work. A modeling framework is developed including a biomechanical model of the knee joint, a finite element model of the knee bearing with encapsulated transducer, and an electromechanical model of the piezoelectric transducer. Model predictions show that a peak voltage of 2.3 V with a load resistance of 1.01 MΩ can be obtained from a single embedded piezoelectric stack, and an average power of 12 μW can be obtained from a knee bearing with four embedded piezoelectric transducers. Uniaxial compression testing is also performed on a fabricated sample for model validation. The results found in this paper show promising potential of embedded piezoelectric transducers to be utilized for autonomous self-powered in vivo knee implant force sensors.Keywords
Funding Information
- National Institutes of Health (R15AR068663)
This publication has 32 references indexed in Scilit:
- Low-power circuits for the wireless ligament balance measurement system in TKAAnalog Integrated Circuits and Signal Processing, 2011
- Patient Satisfaction after Total Knee Arthroplasty: Who is Satisfied and Who is Not?Clinical Orthopaedics and Related Research, 2010
- A review of power harvesting using piezoelectric materials (2003–2006)Smart Materials and Structures, 2007
- Soft tissue balance measurement in anterior cruciate ligament-resected knee joint: cadaveric study as a model for cruciate-retaining total knee arthroplastyJournal of Orthopaedic Science, 2007
- In vivo knee moments and shear after total knee arthroplastyJournal of Biomechanics, 2007
- A multiaxial force-sensing implantable tibial prosthesisJournal of Biomechanics, 2006
- The Use of Piezoelectric Ceramics for Electric Power Generation Within Orthopedic ImplantsIEEE/ASME Transactions on Mechatronics, 2005
- On Low-Frequency Electric Power Generation With PZT CeramicsIEEE/ASME Transactions on Mechatronics, 2005
- Cyclic steady state stress–strain behavior of UHMW polyethyleneBiomaterials, 2000
- Instrumented implant for measuring tibiofemoral forcesJournal of Biomechanics, 1996