Abstract
Objectives To evaluate head CT protocol developed to improve visibility of the brainstem and cerebellum, lower bone-related artefacts in the posterior fossa and maintain patient radioprotection. Methods A paired comparison of head CT performed without Adaptive Statistical Iterative Reconstruction (ASiR) and a clinically indicated follow-up with 40 % ASiR was acquired in one group of 55 patients. Patients were scanned in the axial mode with different scanner settings for the brain and the posterior fossa. Objective image quality analysis was performed with signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Subjective image quality analysis was based on brain structure visibility and evaluation of the artefacts. Results We achieved 19 % reduction of total DLP and significantly better image quality of posterior fossa structures. SNR for white and grey matter in the cerebellum were 34 % to 36 % higher, respectively, CNR was improved by 142 % and subjective analyses were better for images with ASiR. Conclusions When imaging parameters are set independently for the brain and the posterior fossa imaging, ASiR has a great potential to improve CT performance: image quality of the brainstem and cerebellum is improved, and radiation dose for the brain as well as total radiation dose are reduced. Key Points •With ASiR it is possible to lower radiation dose or improve image quality •Sequentional imaging allows setting scan parameters for brain and posterior-fossa independently •We improved visibility of brainstem structures and decreased radiation dose •Total radiation dose (DLP) was decreased by 19 %Keywords
This publication has 20 references indexed in Scilit:
- Recent Advances in CT Image ReconstructionCurrent Radiology Reports, 2013
- Comparison of adaptive statistical iterative and filtered back projection reconstruction techniques in brain CTEuropean Journal of Radiology, 2012
- Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort studyThe Lancet, 2012
- Cranial CT with Adaptive Statistical Iterative Reconstruction: Improved Image Quality with Concomitant Radiation Dose ReductionAmerican Journal of Neuroradiology, 2011
- Lowering the Dose in Head CT Using Adaptive Statistical Iterative ReconstructionAmerican Journal of Neuroradiology, 2011
- Innovations in CT Dose Reduction Strategy: Application of the Adaptive Statistical Iterative Reconstruction AlgorithmAmerican Journal of Roentgenology, 2010
- Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility StudyAmerican Journal of Roentgenology, 2009
- Estimated Risks of Radiation-Induced Fatal Cancer from Pediatric CTAmerican Journal of Roentgenology, 2001
- Thick-Section Reformatting of Thinly Collimated Helical CT for Reduction of Skull Base-Related ArtifactsAmerican Journal of Roentgenology, 2000
- Cranial CT Artifacts and Gantry AngulationJournal of Computer Assisted Tomography, 1991