Precision of glucose measurements in control sera by isotope dilution/mass spectrometry: proposed definitive method compared with a reference method.
Authors:
- O PelletierOPO Pelletier
- C ArratoonCAC Arratoon
Open Access
- 1 August 1987
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 33 (8) , 1397-1402
Abstract
This improved isotope-dilution gas chromatographic/mass spectrometric (GC/MS) method, in which [13C]glucose is the internal standard, meets the requirements of a Definitive Method. In a first study with five reconstituted lyophilized sera, a nested analysis of variance of GC/MS values indicated considerable among-vial variation. The CV for 32 measurements per serum ranged from 0.5 to 0.9%. However, concentration and uncertainty values (mmol/L per gram of serum) assigned to one serum by the NBS Definitive Method (7.56 +/- 0.28) were practically identical to those obtained with the proposed method (7.57 +/- 0.20). In the second study, we used twice more [13C]glucose diluent to assay four serum pools and two lyophilized sera. The CV ranged from 0.26 to 0.5% for the serum pools and from 0.28 to 0.59% for the lyophilized sera. In comparison, results by the hexokinase/glucose-6-phosphate dehydrogenase reference method agreed within acceptable limits with those by the Definitive Method but tended to be slightly higher (up to 3%) for lyophilized serum samples or slightly lower (up to 2.5%) for serum pools.Keywords
This publication has 3 references indexed in Scilit:
- Glass capillary or fused-silica gas chromatography—mass spectrometry of several monosaccharides and related sugars: Improved resolutionJournal of Chromatography B: Biomedical Sciences and Applications, 1982
- Production of low glucose serum for quality control and evaluation of yeast treatment, lyophilization and storage conditions on serum constituentsClinical Biochemistry, 1980
- TOTAL SERUM-CHOLESTEROL BY ISOTOPE DILUTION-MASS SPECTROMETRY - A CANDIDATE DEFINITIVE METHOD1980