MR-compatible assistance system for punction in a high-field system: device and feasibility of transgluteal biopsies of the prostate gland
Authors:
- 10 October 2006
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 17 (4) , 1118-1124
Abstract
We present the first cadavic study results concerning the feasibility of the use of an MR-guided assistance system, Innomotion (Innomedic, Herxheim, Germany), for accurate and consistent placement of percutaneous needles in the prostate gland. The MR-compatible assistance system consists of a C-arch, guiding arm and application module (AMO). T1-weighted fast low angle shot (FLASH) 2-D-GRE sequence (TR/TE=110/4 ms) and T2-weighted turbo spin-echo (TSE)-sequences (TR/TE=3200/97 ms) in transversal orientation were used for the monitoring of the punction of the prostate gland. Planning and control of the intervention is to be made outside the scanner room on a desktop computer that receives DICOM images from the scanner. Servopneumatic drives move the AMO to the insertion point. The physician has to introduce the punction needle manually. The mean deviation of the needle tip to the target in a gel phantom was 0.35 mm. An accurate punction of the prostate gland can easily be performed using this system with a transgluteal access. The T2-weighted images are superior for the evaluation of the prostate anatomy and the needle position during the interventions. In conclusion, our preliminary results indicate that this MR-guided assistance system is suitable for an accurate transgluteal needle placement in the prostate.Keywords
This publication has 41 references indexed in Scilit:
- Cancer Statistics, 2006CA: A Cancer Journal for Clinicians, 2006
- PROSPECTIVE EVALUATION OF PROSTATE CANCER DETECTED ON BIOPSIES 1, 2, 3 AND 4: WHEN SHOULD WE STOP?Journal of Urology, 2001
- PROSTATE CANCER DIAGNOSIS USING A SATURATION NEEDLE BIOPSY TECHNIQUE AFTER PREVIOUS NEGATIVE SEXTANT BIOPSIESJournal of Urology, 2001
- CT-Integrated Robot for Interventional Procedures: Preliminary Experiment and Computer-Human InterfacesComputer Aided Surgery, 2001
- CT-integrated robot for interventional procedures: Preliminary experiment and computer-human interfacesComputer Aided Surgery, 2001
- A PROSPECTIVE RANDOMIZED TRIAL COMPARING 6 VERSUS 12 PROSTATE BIOPSY CORES: IMPACT ON CANCER DETECTIONJournal of Urology, 2000
- Real-Time Magnetic Resonance Imaging-Guided Brachytherapy in the Treatment of Selected Patients with Clinically Localized Prostate CancerJournal of Endourology, 2000
- RISKS AND COMPLICATIONS OF TRANSRECTAL ULTRASOUND GUIDED PROSTATE NEEDLE BIOPSYJournal of Urology, 1998
- The National Cancer Data Base report on prostate carcinoma after the peak in incidence rates in the U. S.Cancer, 1998
- Prostatic biopsy after proctocolectomy: a transgluteal, CT-guided approach.American Journal of Roentgenology, 1996