COMPARISON OF TWO TYPES EYE LENS DOSEMETERS USED IN INTERVENTIONAL CARDIOLOGY AND RADIOLOGY
- 22 August 2022
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 198 (9-11) , 580-585
Abstract
The aim of this contribution is to provide an overview of comparison of two types of eye lens dosemeter systems. The comparison was performed at the Department of Intravenous Cardiology using patient and physician phantoms and supplemented by dose simulation using the Monte Carlo method. The tests were performed in several specific geometries and in addition to eye lens dosemeters the value of a personal dosemeter located at the reference point was also monitored. The value of Hp(3) and Hp(10) was monitored. It is clear from the results that film dosemeters achieve more correct results in most cases. It is probably due to a better correction for the angle of exposure. This assumption must be verified by more detailed measurements in laboratory conditions.Keywords
Funding Information
- European Regional Development Fund-Project (CZ.02.1.01/0.0/0.0/16_019/0000778)
This publication has 6 references indexed in Scilit:
- CONVERSION COEFFICIENTS FROM AIR KERMA TO DIRECTIONAL DOSE EQUIVALENT H′(0.07) FOR PHOTONSRadiation Protection Dosimetry, 2019
- Computational phantoms, ICRP/ICRU, and further developmentsAnnals of the ICRP, 2018
- Air kerma to HP(3) conversion coefficients for photons from 10 keV to 10 MeV, calculated in a cylindrical phantomRadiation Protection Dosimetry, 2012
- Monte Carlo calculations on extremity and eye lens dosimetry for medical staff at interventional radiology proceduresRadiation Protection Dosimetry, 2011
- Comparison of double dosimetry algorithms for estimating the effective dose in occupational dosimetry of interventional radiology staffRadiation Protection Dosimetry, 2008
- Monte Carlo simulations of occupational radiation doses in interventional radiologyBritish Journal of Radiology, 2007