Comparison of dose calculation algorithms in slab phantoms with cortical bone equivalent heterogeneities
- 26 July 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (8) , 3323-3333
Abstract
To evaluate the dose values predicted by several calculation algorithms in two treatment planning systems, Monte Carlo (MC) simulations and measurements by means of various detectors were performed in heterogeneous layer phantoms with water‐ and bone‐equivalent materials. Percentage depth doses (PDDs) were measured with thermoluminescent dosimeters (TLDs), metal‐oxide semiconductor field‐effect transistors (MOSFETs), plane parallel and cylindrical ionization chambers, and beam profiles with films. The MC code used for the simulations was the PENELOPE code. Three different field sizes (, , and ) were studied in two phantom configurations and a bone equivalent material. These two phantom configurations contained heterogeneities of 5 and of bone, respectively. We analyzed the performance of four correction‐based algorithms and one based on convolution superposition. The correction‐based algorithms were the Batho, the Modified Batho, the Equivalent TAR implemented in the Cadplan (Varian) treatment planning system (TPS), and the Helax‐TMS Pencil Beam from the Helax‐TMS (Nucletron) TPS. The convolution‐superposition algorithm was the Collapsed Cone implemented in the Helax‐TMS. All the correction‐based calculation algorithms underestimated the dose inside the bone‐equivalent material for compared to MC simulations. The maximum underestimation, in terms of root‐mean‐square (RMS), was about 15% for the Helax‐TMS Pencil Beam (Helax‐TMS PB) for a field inside the bone‐equivalent material. In contrast, the Collapsed Cone algorithm yielded values around 3%. A more complex behavior was found for where the Collapsed Cone performed less well, overestimating the dose inside the heterogeneity in 3%–5%. The rebuildup in the interface bone‐water and the penumbra shrinking in high‐density media were not predicted by any of the calculation algorithms except the Collapsed Cone, and only the MC simulations matched the experimental values within the estimated uncertainties. The TLD and MOSFET detectors were suitable for dose measurement inside bone‐equivalent materials, while parallel ionization chambers, applying the same calibration and correction factors as in water, systematically underestimated dose by 3%–5%.Keywords
This publication has 45 references indexed in Scilit:
- Dose evaluation in lung-equivalent media in high-energy photon external radiotherapyRadiation Protection Dosimetry, 2006
- Dosimetric verification of a commercial collapsed cone algorithm in simulated clinical situationsRadiotherapy and Oncology, 2004
- Oral Sequelae of Head and Neck RadiotherapyCritical Reviews in Oral Biology & Medicine, 2003
- Pelvic fractures following irradiation of endometrial and vaginal cancers–a case series and review of literatureRadiotherapy and Oncology, 2000
- Characteristics of LiF:Mg,Cu,P for Clinical ApplicationsRadiation Protection Dosimetry, 1999
- Evaluation of lung dose correction methods for photon irradiations of thorax phantomsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1989
- The influence of ionization chamber and phantom design on the measurement of lung dose in photon beamsMedical Physics, 1988
- Dose perturbations at interfaces in photon beamsMedical Physics, 1987
- Accuracy of lung dose calculations for large‐field irradiation with 6‐MV x raysMedical Physics, 1986
- Underdosing of lesions resulting from lack of electronic equilibrium in upper respiratory air cavities irradiated by 10 mv X-ray beamsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1977