Adverse reaction to metal debris after ReCap-M2A-Magnum large-diameter-head metal-on-metal total hip arthroplasty
Open Access
- 31 October 2013
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 84 (6) , 549-554
Abstract
Background and purpose The clinical findings of adverse reaction to metal debris (ARMD) following large-diameter-head metal-on-metal total hip arthroplasty (LDH MoM THA) may include periarticular fluid collections, soft tissue masses, and gluteal muscle necrosis. The ReCap-M2a-Magnum LDH MoM THA was the most commonly used hip device at our institution from 2005 to 2012. We assessed the prevalence of and risk factors for ARMD with this device. Methods 74 patients (80 hips) had a ReCap-M2a-Magnum LDH MoM THA during the period August 2005 to December 2006. These patients were studied with hip MRI, serum chromium and cobalt ion measurements, the Oxford hip score questionnaire, and by clinical examination. The prevalence of ARMD was recorded and risk factors for ARMD were assessed using logistic regression models. The mean follow-up time was 6.0 (5.5–6.7) years. Results A revision operation due to ARMD was needed by 3 of 74 patients (3 of 80 hips). 8 additional patients (8 hips) had definite ARMD, but revision was not performed. 29 patients (32 hips) were considered to have a probable or possible ARMD. Altogether, 43 of 80 hips had a definite, probable, or possible ARMD and 34 patients (37 hips) were considered not to have ARMD. In 46 of 78 hips, MRI revealed a soft tissue mass or a collection of fluid (of any size). The symptoms clicking in the hip, local hip swelling, and a feeling of subluxation were associated with ARMD. Interpretation ARMD is common after ReCap-M2a-Magnum total hip arthroplasty, and we discourage the use of this device. Asymptomatic patients with a small fluid collection on MRI may not need instant revision surgery but must be followed up closely.Keywords
This publication has 24 references indexed in Scilit:
- Metal Ion Levels Not Sufficient as a Screening Measure for Adverse Reactions in Metal-on-Metal Hip ArthroplastiesThe Journal of Arthroplasty, 2013
- Hip resurfacing in a district general hospital: 6-year clinical results using the ReCap hip resurfacing systemBMC Musculoskeletal Disorders, 2012
- One-year prospective comparative study of three large-diameter metal-on-metal total hip prostheses: Serum metal ion levels and clinical outcomesOrthopaedics & Traumatology: Surgery & Research, 2012
- Low Incidence of Groin Pain and Early Failure with Large Metal Articulation Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2012
- Noninflammatory Pseudotumor Simulating Venous Thrombosis After Metal-on-Metal Hip ResurfacingOrthopedics, 2011
- Metal Ion Release With Large-Diameter Metal-on-Metal Hip ArthroplastyThe Journal of Arthroplasty, 2011
- Analysis of wear of retrieved metal-on-metal hip resurfacing implants revised due to pseudotumoursThe Journal of Bone and Joint Surgery. British volume, 2010
- Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2010
- Early clinical failure of the Birmingham metal-on-metal hip resurfacing is associated with metallosis and soft-tissue necrosisThe Journal of Bone and Joint Surgery. British volume, 2009
- Differences in the fretting corrosion of metal–metal and ceramic–metal modular junctions of total hip replacementsJournal of Orthopaedic Research, 2004