Non-Contact Skin Moisture Measurement Based on Near-Infrared Spectroscopy
- 1 December 2004
- journal article
- clinical trial
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 58 (12) , 1439-1446
Abstract
Non-contact skin moisture measurement based on near-infrared (NIR) spectroscopy is proposed in the spectral range from 1300 to 2000 nm. A gap is introduced between the optical fiber probe and the skin surface in order to avoid occluding surface vapor. In vitro and in vivo experiments for measuring the water content of skin are implemented. The measured absorbance spectra are processed by multivariate analyses. Processed results are compared with the water content values obtained by a capacitance method. The correlations between the optical method and the capacitance method obtained by partial-least squares regression are higher than those obtained by multiple linear regression. In addition, a Monte Carlo simulation is implemented to evaluate measurement depths of the optical methods. It is presented that the measurement depth of the optical method depends largely on the water absorption. The simulation result also shows that the measurement depth of the optical method is much deeper than the depth of the capacitance method, especially in the spectral range where water absorption is relatively weak.Keywords
This publication has 16 references indexed in Scilit:
- Distinct locational differences observable in biophysical functions of the facial skin: with special emphasis on the poor functional properties of the stratum corneum of the perioral regionInternational Journal of Cosmetic Science, 2004
- Monte Carlo Simulation of Near Infrared Reflectance Spectroscopy in the Wavelength Range from 1000 nm to 1900 nmOptical Review, 2003
- Temperature-Insensitive Measurement of Glucose Concentration Based on Near Infrared Spectroscopy and Partial Least Squares AnalysisOptical Review, 2003
- Temperature modulation of the visible and near infrared absorption and scattering coefficients of human skinJournal of Biomedical Optics, 2003
- Long-wavelength near-infrared spectroscopic imaging for in-vivo skin hydration measurementsVibrational Spectroscopy, 2002
- In vivo Measurements of Water in Skin by Near-Infrared ReflectanceApplied Spectroscopy, 1998
- EEMCO guidance for the assessment of stratum corneum hydration: electrical methodsSkin Research and Technology, 1997
- Light scattering in lntralipid-10% in the wavelength range of 400–1100 nmApplied Optics, 1991
- A review of the optical properties of biological tissuesIEEE Journal of Quantum Electronics, 1990
- Stratum corneum hydration and amino acid content in xerotic skinBritish Journal of Dermatology, 1989