Evaluation of the monocyte counting by two automated haematology analysers compared with flow cytometry
- 1 April 2005
- journal article
- research article
- Published by Wiley in Clinical and Laboratory Haematology
- Vol. 27 (2) , 91-97
Abstract
The aim is to determine the monocyte count performance of the Bayer Diagnostics ADVIA120 and Coulter LH 750 automated haematology analysers and the results obtained by these two instruments were compared with those provided by Becton Dickinson FACScan flow cytometer using the combination of CD45/CD14 MoAb. Linearity and imprecision were also evaluated. The linearity of both instruments was good. Coulter LH 750 showed better precision (4.3%) than ADVIA 120 (9.0%) both within and between batch. A significant correlation (r = 0.973) was found between the LH 750 and the flow cytometry method, while a modest one was observed between the latter and the ADVIA 120 (r = 0.880). When comparing the percentage of monocytes by means of one-way anova and Tukey test, it was found that the LH 750 provided the closest results in comparison with flow cytometry, with no statistical difference between the means (mean difference MO% = 0.6); however the difference was statistically different between the ADVIA 120 and flow cytometry (mean difference MO% = -4.06). These data were confirmed by Altman-Bland and Deming regression analyses.Keywords
This publication has 14 references indexed in Scilit:
- Evaluation of the Abbott CELL-DYN 4000 Hematology AnalyzerAmerican Journal of Clinical Pathology, 2000
- Analyses of quality assessment studies using CD45 for gating lymphocytes for CD3+4+%Cytometry, 2000
- Evaluation of Four Automated Hematology Analyzers: A Comparative Study of Differential Counts (Imprecision and Inaccuracy)American Journal of Clinical Pathology, 1992
- Monocyte counting: discrepancies in results obtained with different automated instruments.Journal of Clinical Pathology, 1991
- The Author’s Reply Analytic Goals Are Targets, Not Inflexible Criteria of AcceptabilityAmerican Journal of Clinical Pathology, 1988
- Special Report: Desirable Standards for Hematology Tests: A ProposalAmerican Journal of Clinical Pathology, 1987
- The Author’s Reply: Thanks, Appreciation, and CelebrationAmerican Journal of Clinical Pathology, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Proposals for the classification of the myelodysplastic syndromesBritish Journal of Haematology, 1982
- Simple and rapid measurement of human T lymphocytes and their subclasses in peripheral blood.Proceedings of the National Academy of Sciences, 1980