Model‐based crosstalk compensation for simultaneous dual‐isotope brain SPECT imaging
Authors:
- Yong DuYDYong Du
- Benjamin M. W. TsuiBTBenjamin M. W. Tsui
- Eric C. FreyEFEric C. Frey
- 14 August 2007
- journal article
- Published by Wiley in Medical Physics
- Vol. 34 (9) , 3530-3543
Abstract
In this work, we developed a model‐based method to estimate and compensate for the crosstalk contamination in simultaneous and dual isotope brain single photo emission computed tomography imaging. The model‐based crosstalk compensation (MBCC) includes detailed modeling of photon interactions inside both the object and the detector system. In the method, scatter in the object is modeled using the effective source scatter estimation technique, including contributions from all the photon emissions. The effects of the collimator‐detector response, including the penetration and scatter components due to high‐energy photons, are modeled using pre‐calculated tables of Monte Carlo simulated point‐source response functions obtained from sources in air at various distances from the face of the collimator. The model‐based crosstalk estimation method was combined with iterative reconstruction based compensation to reduce contamination due to crosstalk. The MBCC method was evaluated using Monte Carlo simulated and physical phantom experimentally acquired simultaneous dual‐isotope data. Results showed that, for both experimental and simulation studies, the model‐based method provided crosstalk estimates that were in good agreement with the true crosstalk. Compensation using MBCC improved image contrast and removed the artifacts for both Monte Carlo simulated and experimentally acquired data. The results were in good agreement with images acquired without any crosstalk contamination.Keywords
Funding Information
- Public Health Service (R01-EB00288)
This publication has 23 references indexed in Scilit:
- Quantification in simultaneous99mTc/123I brain SPECT using generalized spectral factor analysis: a Monte Carlo studyPhysics in Medicine & Biology, 2006
- Model-based compensation for quantitative123I brain SPECT imagingPhysics in Medicine & Biology, 2006
- Optimization of acquisition energy windows in simultaneous /sup 99m/Tc//sup 123/I brain SPECTIEEE Transactions on Nuclear Science, 2003
- Radiopharmaceuticals for single-photon emission computed tomography brain imagingSeminars in Nuclear Medicine, 2003
- Combination of MCNP and SimSET for Monte Carlo simulation of SPECT with medium- and high-energy photonsIEEE Transactions on Nuclear Science, 2002
- Scatter and cross-talk corrections in simultaneous Tc-99m/I-123 brain SPECT using constrained factor analysis and artificial neural networksIEEE Transactions on Nuclear Science, 2000
- Clinical value of triple-energy window scatter correction in simultaneous dual-isotope single-photon emission tomography with123I-BMIPP and201TIEuropean Journal of Nuclear Medicine and Molecular Imaging, 1997
- A vector Wiener filter for dual-radionuclide imagingIEEE Transactions on Medical Imaging, 1996
- Feasibility of dual radionuclide brain imaging with I‐123 and Tc‐99 mMedical Physics, 1994
- Three-dimensional iterative reconstruction algorithms with attenuation and geometric point response correctionIEEE Transactions on Nuclear Science, 1991