Effect of left ventricular assist device on the hemodynamics of a patient-specific left heart
Authors:
- 20 April 2022
- journal article
- research article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 60 (6) , 1705-1721
Abstract
This article describes the numerical efforts made to investigate the influence of a left ventricular assist device (LVAD) on the patient-specific left heart’s hemodynamics. Two different computational geometries with left heart have been simulated over the entire cardiac cycle (case 1: healthy heart without LVAD and case 2: diseased heart with LVAD). The blood flow was simulated by implementing Bird-Carreau non-Newtonian model. Simulation results show that implantation of LVAD pump imparts major influence on the hemodynamics of the heart; it also provides a cardiac output of 4.87 L/min even at the diastolic phase. Furthermore, post LVAD implantation, approximately eight times more wall shear stress, is noticed at the aorta during the ventricular systole. In particular, major changes in the fluidics are observed inside the aortic region. A possibility of flow stagnation is noticed near the aortic root during the diastolic phase due to the bisection of incoming bloodstreams from the outflow graft. The flow characteristics of the LVAD pump are also observed to be significantly different from the idealized simulations (idealized tubular inlet situation). The observation of this study can help in understanding post-implant critical hemodynamic issues due to pump performance and its subsequent impact on the heart. Graphical abstract A simulation approach-based study has been performed to investigate the influence of LVAD on the hemodynamics of a heart. A 3D computational model of a patient-specific heart has been created from CT scan datasets for diastole and systole phases (a). An axial flow blood pump has been implanted computationally into the left heart (b). The implanted blood pump enhances the cardiac output and elevates shear generation (c) and (d).Keywords
This publication has 47 references indexed in Scilit:
- The Global Health and Economic Burden of Hospitalizations for Heart FailureJACC, 2014
- Computational fluid dynamics in patients with continuous-flow left ventricular assist device support show hemodynamic alterations in the ascending aortaThe Journal of Thoracic and Cardiovascular Surgery, 2014
- Impact of Pulmonary Venous Locations on the Intra-Atrial Flow and the Mitral Valve Plane Velocity ProfileCardiovascular Engineering and Technology, 2012
- Ascending aorta backward flow parameters estimated from phase-contrast cardiovascular magnetic resonance data: new indices of arterial agingJournal of Cardiovascular Magnetic Resonance, 2012
- Computational Characterization of Flow and Hemolytic Performance of the UltraMag Blood Pump for Circulatory SupportArtificial Organs, 2010
- Fluid–structure interaction analysis of a patient‐specific right coronary artery with physiological velocity and pressure waveformsCommunications in Numerical Methods in Engineering, 2009
- Non-Newtonian blood flow in human right coronary arteries: steady state simulationsJournal of Biomechanics, 2004
- Effects of acute left ventricular unloading on right ventricular function in normal and chronic right ventricular pressure-overloaded lambsThe Journal of Thoracic and Cardiovascular Surgery, 2003
- Implantable Left Ventricular Assist DevicesNew England Journal of Medicine, 1998
- Red Cell Injury Assessed in a Numeric Model of a Peripheral Dialysis NeedleASAIO Journal, 1996