Electrodeposition of chitosan-glucose oxidase biocomposite onto Pt–Pb nanoparticles modified stainless steel needle electrode for amperometric glucose biosensor
Authors:
- 14 June 2011
- journal article
- Published by Springer Nature in Journal of Materials Science: Materials in Medicine
- Vol. 22 (8) , 1985-1992
Abstract
A glucose biosensor was fabricated by electrodepositing chitosan (CS)-glucose oxidase(GOD) biocomposite onto the stainless steel needle electrode (SSN electrode) modified by Pt–Pb nanoparticles (Pt–Pb/SSN electrode). Firstly, Pt–Pb nanoparticles were deposited onto the SSN electrode and then CS-GOD biocomposite was co-electrodeposited onto the Pt–Pb/SSN electrode in a mixed solution containing p-benzoquinone (p-BQ), CS and GOD. The electrochemical results showed that the Pt–Pb nanoparticles can accelerate the electron transfer and improve the effective surface area of the SSN electrode. As a result, the detection range of the proposed biosensor was from 0.03 to 9 mM with a current sensitivity of 0.4485 μA/mM and a response time of 15 s. The Michaelis constant value was calculated to be 4.9837 mM. The cell test results indicated that the electrodes have a low cytotoxicity. This work provided a suitable technology for the fabrication of the needle-type glucose biosensor.Keywords
This publication has 34 references indexed in Scilit:
- Corrigendum to “AC-electrophoretic deposition of glucose oxidase” [Biosens. Bioelectron. 25 (1) (2009) 191–197]Biosensors and Bioelectronics, 2010
- AC-electrophoretic deposition of glucose oxidaseBiosensors and Bioelectronics, 2009
- Fluorescent Si nanoparticle-based electrode for sensing biomedical substancesOptics Communications, 2008
- Electrodeposition of Carbon Nanotubes−Chitosan−Glucose Oxidase Biosensing Composite Films Triggered by Reduction of p-Benzoquinone or H2O2The Journal of Physical Chemistry B, 2007
- An iridium nanoparticles dispersed carbon based thick film electrochemical biosensor and its application for a single use, disposable glucose biosensorSensors and Actuators B: Chemical, 2007
- Enhanced amperometric detection of glucose using Si29 particlesApplied Physics Letters, 2006
- A novel method for glucose determination based on electrochemical impedance spectroscopy using glucose oxidase self-assembled biosensorBioelectrochemistry, 2006
- Development of a highly responsive needle-type glucose sensor using polyimide for a wearable artificial endocrine pancreasJournal of Artificial Organs, 2006
- Tissue implanted glucose needle electrodes: early sensor stabilisation and achievement of tissue-blood correlation during the run in periodAnalytica Chimica Acta, 2005
- Signal-Directed Sequential Assembly of Biomolecules on Patterned SurfacesLangmuir, 2005