Ion Suppression Study for Tetracyclines in Feed
Authors:
Open Access
- 27 September 2012
- journal article
- research article
- Published by Hindawi Limited in Chromatography Research International
- Vol. 2012, 1-9
Abstract
Ion suppression in analysis of tetracyclines in feed was studied. The conventional analysis consists of a liquid extraction followed by a clean-up step using solid phase extraction (SPE) technique and analysis of the tetracyclines by liquid chromatography and mass spectrometric detection. Various strategies for extraction and cleanup were tested in the present work, and the effectiveness to decrease the ion suppression on the MS/MS signals was evaluated. Four sample treatment methods were tested with five different feed samples. Extraction solvents tested were McIlvaine buffer and a mixture of McIlvaine buffer dichloromethane (3 : 1). SPE cartridges for cleanup were Oasis HLB, Oasis MCX, and Oasis MAX. The effectiveness of the methods was evaluated in terms of decreasing the ion suppression effect but also of decreasing the variability of ion suppression between samples. The method that provided the most satisfactory results involved a clean-up step based on SPE using mixed-mode cation exchange cartridges (Oasis MCX).Keywords
This publication has 29 references indexed in Scilit:
- Complexities in tetracycline analysis—chemistry, matrix extraction, cleanup, and liquid chromatographyJournal of Chromatography A, 2005
- The ion suppression phenomenon in liquid chromatography–mass spectrometry and its consequences in the field of residue analysisAnalytica Chimica Acta, 2005
- A study of ion suppression effects in electrospray ionization from mobile phase additives and solid‐phase extractsRapid Communications in Mass Spectrometry, 2003
- Ion Suppression in Mass SpectrometryClinical Chemistry, 2003
- Strategies for the Assessment of Matrix Effect in Quantitative Bioanalytical Methods Based on HPLC−MS/MSAnalytical Chemistry, 2003
- Contamination of animal feedingstuffs as a cause of residues in food: a review of regulatory aspects, incidence and controlAnalytica Chimica Acta, 2002
- Ion suppression effects in liquid chromatography–electrospray-ionisation transport-region collision induced dissociation mass spectrometry with different serum extraction methods for systematic toxicological analysis with mass spectra librariesJournal of Chromatography B, 2002
- Quantitative screening and matrix effect studies of drug discovery compounds in monkey plasma using fast‐gradient liquid chromatography/tandem mass spectrometryRapid Communications in Mass Spectrometry, 2001
- Chromatographic analysis of tetracycline antibiotics in foodsJournal of Chromatography A, 2000
- The effects of sample preparation methods on the variability of the electrospray ionization response for model drug compoundsRapid Communications in Mass Spectrometry, 1999