Scanner and kVp dependence of measured CT numbers in the ACR CT phantom
Authors:
Open Access
- 4 November 2013
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 14 (6) , 338-349
Abstract
Quality control testing of CT scanners in our region includes a measurement of CT numbers in the American College of Radiology (ACR) CT phantom using a standardized protocol. CT number values are clinically relevant in determining the composition of various tissues in the body. Accuracy is important in the characterization of tumors, assessment of coronary calcium, and identification of urinary stone composition. Effective quality control requires that tolerance ranges of CT number values be defined: a measured value outside the range indicates the need for further investigation and possible recalibration of the scanner. This paper presents the results of CT number measurements on 36 scanners (25 GE, 10 Siemens and 1 Toshiba) at each available kVp. Among the five materials (solid water, air, polyethylene, acrylic, bone-equivalent) the measured CT numbers exhibit manufacturer and kVp dependence, which should be taken into account when defining tolerances. With this scan protocol, air and solid water values are significantly higher on GE scanners than on Siemens scanners ( at each kVp). The CT numbers of polyethylene and acrylic increase with kVp, while the bone-equivalent CT number decreases. These results are used to define manufacturer- and kVp-specific tolerance ranges for the CT numbers of each material in this phantom, which will be used in our quality control program. PACS numbers: 87.57.qp, 87.57.C-
Keywords
This publication has 15 references indexed in Scilit:
- Pulmonary CT Angiography as First-Line Imaging for PE: Image Quality and Radiation Dose ConsiderationsAmerican Journal of Roentgenology, 2013
- Radiologic and Nuclear Medicine Studies in the United States and Worldwide: Frequency, Radiation Dose, and Comparison with Other Radiation Sources—1950–2007Radiology, 2009
- Application‐ and patient size‐dependent optimization of x‐ray spectra for CTMedical Physics, 2009
- Population exposure to ionizing radiation from medical examinations in FranceBritish Journal of Radiology, 2008
- Measuring Coronary Calcium on CT Images Adjusted for Attenuation DifferencesRadiology, 2005
- The phantom portion of the American College of Radiology (ACR) Computed Tomography (CT) accreditation program: Practical tips, artifact examples, and pitfalls to avoidMedical Physics, 2004
- Adrenal Masses: Characterization with Combined Unenhanced and Delayed Enhanced CTRadiology, 2002
- Age and gender distributions of coronary artery calcium detected by electron beam tomography in 35,246 adultsThe American Journal of Cardiology, 2001
- An algorithm for noninvasive identification of angiographic three-vessel and/or left main coronary artery disease in symptomatic patients on the basis of cardiac risk and electron-beam computed tomographic calcium scoresJACC, 1999
- Quantification of coronary artery calcium using ultrafast computed tomographyJACC, 1990