Validity of GNRB® arthrometer compared to Telos™ in the assessment of partial anterior cruciate ligament tears
- 22 January 2013
- journal article
- research article
- Published by Wiley in Knee Surgery, Sports Traumatology, Arthroscopy
- Vol. 22 (2) , 285-290
Abstract
Purpose: The main goal of this study was to compare the results of the GNRB® arthrometer to those of Telos™ in the diagnosis of partial thickness tears of the anterior cruciate ligament (ACL).Methods: A prospective study performed January–December 2011 included all patients presenting with a partial or full‐thickness ACL tears without ACL reconstruction and with a healthy contralateral knee. Anterior laxity was measured in all patients by the Telos™ and GNRB® devices. This series included 139 patients, mean age 30.7 ± 9.3 years. Arthroscopic reconstruction was performed in 109 patients, 97 for complete tears and 12 single bundle reconstructions for partial thickness tears. Conservative treatment was proposed in 30 patients with a partial thickness tear. The correlation between the two devices was evaluated by the Spearman coefficient. The optimal laxity thresholds were determined with ROC curves, and the diagnostic value of the tests was assessed by the area under the curve (AUC).Results: The differential laxities of full and partial thickness tears were significantly different with the two tests. The correlation between the results of laxity measurement with the two devices was fair, with the strongest correlation between Telos™ 250 N and GNRB® 250 N (r = 0.46, p = 0.00001). Evaluation of the AUC showed that the informative value of all tests was fair with the best results with the GNRB® 250 N: AUC = 0.89 [95 % CI 0.83–0.94]. The optimal differential laxity threshold with the GNRB® 250 N was 2.5 mm (Se = 84 %, Sp = 81 %).Conclusion: The diagnostic value of GNRB® was better than Telos™ for ACL partial thickness tears.Level of evidence: Diagnostic study, Level II.Keywords
This publication has 26 references indexed in Scilit:
- Comparison of 5 Different Methods for Measuring Stress Radiographs to Improve Reproducibility During the Evaluation of Knee InstabilityThe American Journal of Sports Medicine, 2011
- Current concept of partial anterior cruciate ligament rupturesOrthopaedics & Traumatology: Surgery & Research, 2010
- Instrumented Examination of Anterior Cruciate Ligament Injuries: Minimizing Flaws of the Manual Clinical ExaminationArthroscopy, 2010
- Current Concepts in Instrumented Knee-Laxity TestingThe American Journal of Sports Medicine, 2008
- A New Mechanical Testing Device for Measuring Anteroposterior Knee LaxityThe American Journal of Sports Medicine, 2004
- Intratester and Intertester Reliability of the KT-1000 Arthrometer in the Assessment of Posterior Laxity of the KneeThe American Journal of Sports Medicine, 1997
- Anterior knee motion analysisThe American Journal of Sports Medicine, 1991
- The natural history of conservatively treated partial anterior cruciate ligament tearsThe American Journal of Sports Medicine, 1989
- Measuring the Accuracy of Diagnostic SystemsScience, 1988
- Instrumented measurement of anterior knee laxity in patients with acute anterior cruciate ligament disruptionThe American Journal of Sports Medicine, 1985