Abstract
The aim of our study is to evaluate clinical long-term results and determine changes in periprosthetic bone density of the custom-made hip prostheses Evolution-K (R) and Adaptiva (R). Periprosthetic bone density were evaluated by means of DEXA (Lunar-iDXA-Prodigy (R) bone densitometer) with a long-term follow-up of 16 (15-18) years (Evolution-K (R)) in 24 patients and 13 (13-15) years (Adaptiva (R)) in 41 patients. Evolution-K (R) had a survival rate of 92% and yielded 79/100 points in Harris Hip Score, a mediocre result. Adaptiva (R) had a survival rate of 99% and achieved a good score of 88/100 points. Bone density measurements demonstrated the greatest loss of bone density in the proximal regions of interest (ROI) for both prosthesis types (Evolution-K (R): -25.8% ROI 1, -40.3% ROI 7; -8.3% ROI 2, -10.4% ROI 6; Adaptiva (R): -29.8% ROI 7, -6.8% ROI 6, +14.3% ROI 3, +3.1% ROI 4). Adaptiva (R) yielded a good clinical result as compared to Evolution-K (R) with only average clinical results. Both prostheses clearly showed signs of "stress shielding". Here, the Adaptiva (R) achieved reduced bone density loss as compared to the Evolution-K (R)Keywords
This publication has 24 references indexed in Scilit:
- Microsection analysis of cortical form-fit of a custom femoral component in total hip arthroplasty: an in vitro studyOrthopedic Reviews, 2012
- Low BMD affects initial stability and delays stem osseointegration in cementless total hip arthroplasty in womenActa Orthopaedica, 2012
- Custom Cementless THA in Patients with Skeletal Dysplasia Results in Lower Apparent Revision Rates than Other Types of Femoral FixationClinical Orthopaedics and Related Research, 2011
- Uncemented custom femoral components in hip arthroplastyActa Orthopaedica, 2010
- Custom Cementless Stem Improves Hip Function in Young Patients at 15-year FollowupClinical Orthopaedics and Related Research, 2010
- Durability of Second-generation Extensively Porous-coated Stems in Patients Age 50 and YoungerClinical Orthopaedics and Related Research, 2010
- Zementfreie Hüftendoprothetik: eine aktuelle Übersicht / Cementless total hip arthroplasty: a reviewBiomedical Engineering / Biomedizinische Technik, 2010
- Custom stems for femoral deformity in patients less than 40 years of ageActa Orthopaedica, 2009
- Medium- and long-term performance of 11 516 uncemented primary femoral stems from the Norwegian arthroplasty registerThe Journal of Bone and Joint Surgery. British volume, 2007
- Age-Related Decline in Bone Mass Measured by Dual-Energy X-Ray Absorptiometry and Quantitative Ultrasound in a Population-Based Sample of Both SexesJournal of Clinical Densitometry, 2005