Abstract
An autosampler was developed to interface with a gas chromatograph for N 2 O analyses. The sampler was comprised of a modified fraction collector and pneumatic valves all controlled by a microprocessor unit. Cost of the autosampler system including the fraction collector was estimated to be less than $3500. The sampler was designed to analyze gas samples stored in rubber‐stoppered, glass vials. Use of vials resulted in a slight loss of sensitivity due to dilution and absorption of N 2 O; however, accuracy was the same as could be obtained by manual syringe injections.