Amplitude and phase calibration of hydrophones up to 70 MHz using broadband pulse excitation and an optical reference hydrophone
Authors:
- 1 June 2004
- journal article
- conference paper
- Published by Acoustical Society of America (ASA) in The Journal of the Acoustical Society of America
- Vol. 115 (6) , 2892-2903
Abstract
A substitution calibration technique for piezoelectric ultrasonic hydrophones is presented that uses an optical multilayer hydrophone as the reference receiver. Broadband nonlinearly distorted focused pulses are first measured with the reference hydrophone and then with the hydrophone to be calibrated. By Fourier transformation of the time wave forms and division of the frequency spectra, the complex-valued frequency response of the hydrophone under test is obtained in a broad frequency range in a very fast and efficient way and with high frequency resolution. The results obtained for a membrane hydrophone and a needle-type hydrophone are compared with those obtained by independent calibration techniques such as primary calibration using optical interferometry and secondary calibration using time-delay spectrometry, and good agreement is found. The calibration data obtained are apt to improve the results of ultrasound exposure measurements using broadband voltage-to-pressure conversion. This is demonstrated for standard pulse parameter determination from exemplar exposure measurements on a commercial diagnostic ultrasound machine. For the membrane hydrophone, the evaluation method commonly used leads to an overestimation of the positive peak pressure by up to 50%, an underestimation of the rarefactional peak pressure by up to 11%, and an overestimation of the pulse intensity integral by up to 28%.Keywords
This publication has 29 references indexed in Scilit:
- Primary calibration of hydrophones with extended frequency range 1 to 70 MHz using optical interferometryIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1999
- Are current hydrophone low frequency response standards acceptable for measuring mechanical/cavitation indices?Ultrasonics, 1996
- Pressure pulse distortion by hydrophones due to diminished low frequency responseIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1995
- Frequency response of PVDF needle-type hydrophonesUltrasound in Medicine & Biology, 1994
- Study of surface elastic wave induced on backing material and diffracted field of a piezoelectric polymer film hydrophoneThe Journal of the Acoustical Society of America, 1993
- The influence of spatial polarization distribution on spot poled PVDF membrane hydrophone performanceUltrasound in Medicine & Biology, 1992
- A model of the effects of hydrophone and amplifier frequency response on ultrasound exposure measurementsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1991
- Calibration of hydrophones based on reciprocity and time delay spectrometryIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1988
- Primary calibration of ultrasonic hydrophone using optical interferometryIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1988
- Characteristics of a PVDF Membrane Hydrophone for Use in the Range 1-100 MHzIEEE Transactions on Sonics and Ultrasonics, 1982