Robust optimization in IMPT using quadratic objective functions to account for the minimum MU constraint
Authors:
- 17 November 2017
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 45 (1) , 460-469
Abstract
Purpose: Currently, in clinical practice of intensity‐modulated proton therapy (IMPT), the influence of the minimum monitor unit (MU) constraint is taken into account through postprocessing after the optimization is completed. This may degrade the plan quality and plan robustness. This study aims to mitigate the impact of the minimum MU constraint directly during the plan robust optimization.Methods and materials: Cao et al. have demonstrated a two‐stage method to account for the minimum MU constraint using linear programming without the impact of uncertainties considered. In this study, we took the minimum MU constraint into consideration using quadratic optimization and simultaneously had the impact of uncertainties considered using robust optimization. We evaluated our method using seven cancer patients with different machine settings.Result: The new method achieved better plan quality than the conventional method. The D95% of the clinical target volume (CTV) normalized to the prescription dose was (mean [min–max]): (99.4% [99.2%–99.6%]) vs. (99.2% [98.6%–99.6%]). Plan robustness derived from these two methods was comparable. For all seven patients, the CTV dose–volume histogram band gap (narrower band gap means more robust plans) at D95% normalized to the prescription dose was (mean [min–max]): (1.5% [0.5%–4.3%]) vs. (1.2% [0.6%–3.8%]).Conclusion: Our new method of incorporating the minimum MU constraint directly into the plan robust optimization can produce machine‐deliverable plans with better tumor coverage while maintaining high‐plan robustness.Keywords
Funding Information
- National Cancer Institute (K25CA168984)
- Fraternal Order of Eagles
This publication has 51 references indexed in Scilit:
- PTV-based IMPT optimization incorporating planning risk volumes vs robust optimizationMedical Physics, 2013
- Influence of robust optimization in intensity‐modulated proton therapy with different dose delivery techniquesMedical Physics, 2012
- Robust optimization of intensity modulated proton therapyMedical Physics, 2012
- Including robustness in multi-criteria optimization for intensity-modulated proton therapyPhysics in Medicine & Biology, 2012
- Proton Stereotactic Body Radiation Therapy for Clinically Challenging Cases of Centrally and Superiorly Located Stage I Non-Small-Cell Lung CancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Minimax optimization for handling range and setup uncertainties in proton therapyMedical Physics, 2011
- Direct aperture optimization as a means of reducing the complexity of intensity modulated radiation therapy plansRadiation Oncology, 2009
- Reducing the sensitivity of IMPT treatment plans to setup errors and range uncertainties via probabilistic treatment planningMedical Physics, 2008
- Worst case optimization: a method to account for uncertainties in the optimization of intensity modulated proton therapyPhysics in Medicine & Biology, 2008
- Intensity modulated proton therapy and its sensitivity to treatment uncertainties 2: the potential effects of inter-fraction and inter-field motionsPhysics in Medicine & Biology, 2008