Biologic Responses to Orthopedic Implants
Author:
- 1 April 2016
- journal article
- Published by Wolters Kluwer Health in Spine
- Vol. 41, S30-S31
Abstract
Almost 20% of joint replacement implants fail at 15 to 20 years. Reports suggest that systemic effects of metal-on-metal implants and local effects of total joint arthroplasty implants contributing to implant failure are immune system based. Sometimes implant wear debris can cause implant failure resulting from bone fracture, infection, or implant fracture/failure; most often, aseptic osteolysis or loosening leads to wear debris. Debris is produced by wear (primary) or by corrosion. Corrosion-chemical oxidation comprising reduction reactions involving electron transport-produces electrochemical degradation. Metallic implant degradation occurs when electrochemical dissolution and mechanical/physical wear are combined (i.e., tribocorrosion). With metal-on-metal implants, even with relatively low levels of wear and particle release, pathology caused by metal debris such as pseudotumor/fibrous tissue growth can lead to early implant failure.Keywords
This publication has 4 references indexed in Scilit:
- Tribocorrosion and oral and maxillofacial surgical devicesBritish Journal of Oral and Maxillofacial Surgery, 2014
- Fretting crevice corrosion of stainless steel stem–CoCr femoral head connections: Comparisons of materials, initial moisture, and offset lengthJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008
- The Biology of Aseptic OsteolysisClinical Orthopaedics and Related Research, 2007
- Revision hip surgery in the elderlyThe Journal of Arthroplasty, 1999