A new technique for measuring cardiac output and shunt fraction during venovenous extracorporeal membrane oxygenation
Authors:
- 1 January 1999
- journal article
- research article
- Published by SAGE Publications in Perfusion
- Vol. 14 (1) , 43-47
Abstract
A new indicator dilution technique is described for measuring cardiac output and shunt fraction in patients undergoing venovenous extracorporeal membrane oxygenation (ECMO). Shunt fraction is the proportion of the ECMO pump flow which recirculates through the ECMO circuit (passing directly from the inflow cannula to the outflow cannula) instead of flowing through the pulmonary and systemic circulations. The indicator is an isotonic (150 mmol/l) solution of lithium chloride which is injected into the ECMO flow returning to the patient. Two lithium sensors are used simultaneously to record the resulting lithium dilution curves in arterial blood and in the blood in the ECMO circuit. Cardiac output and shunt fraction are derived from these curves. The technique, which is simple and safe, provides measurements that allow optimal adjustment of ECMO flow and cardiovascular support.Keywords
This publication has 12 references indexed in Scilit:
- Acute lung injury and the acute respiratory distress syndromeCritical Care Medicine, 1998
- In vitro evaluation of a new lithium dilution method of measuring cardiac output and shunt fraction in patients undergoing venovenous extracorporeal membrane oxygenationCritical Care Medicine, 1998
- Adult extracorporeal membrane oxygenation for severe respiratory failurePerfusion, 1998
- Lithium dilution cardiac output measurementCritical Care Medicine, 1997
- Extracorporeal Life Support for 100 Adult Patients With Severe Respiratory FailureAnnals of Surgery, 1997
- The Effectiveness of Right Heart Catheterization in the Initial Care of Critically III PatientsJAMA, 1996
- Thermodilution Cardiac Output May Be Incorrect in Patients on Venovenous Extracorporeal Lung AssistAmerican Journal of Respiratory and Critical Care Medicine, 1995
- Cardiac output measurement in critically ill patients: comparison of continuous and conventional thermodilution techniquesCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1995
- Improved oxygenation with reduced recirculation during venovenous extracorporeal membrane oxygenationCritical Care Medicine, 1995
- Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patientsCritical Care Medicine, 1995