Evaluation of an automated microbiologic blood culture device for detection of bacteria in platelet components
Authors:
- 1 July 1998
- journal article
- research article
- Published by Wiley in Transfusion
- Vol. 38 (7) , 674-679
Abstract
BACKGROUND:Automated culture methods have been used by several investigators to detect bacterial contamination of cellular blood components. We investigated several factors affecting detection by automated culture of bacteria in platelet concentrates (PCs).These factors included the initial contamination level in PCs, the PC sample volume, the PC sample time, and the white cell level in relation to bacteria levels in the PCs.STUDY DESIGN AND METHODS:Staphylococcus epidermidis or Escherichia coli was inoculated into freshly prepared PCs or white cell‐reduced PCs to yield colony‐forming unit (CFU) levels of 10, 1, or 0.1 per mL. At the time of inoculation (t=0) and at t=6, t=24, and t=48 hours, 0.5, 1.0, and 2.0 mL samples of the contaminated PCs were transferred into culture bottles. The presence of bacteria in the culture bottles was subsequently monitored by an automated blood culturing instrument. Bacteria levels in the PC at the time of first automated culture detection were determined by quantitative plating.RESULTS:E. coli was detected in 92 percent of experiments when 1.0‐ or 2.0‐mL samples were taken at t=6 hours. At t=24 hours, 100‐percent detection was observed with all tested inoculation volumes; however, by that time,>10(7) CFU per mL of bacteria were present in every PC. For S. epidermidis, 89 percent and 83 percent of contaminated PCs were detected with a t=24 hour sampling time and 2.0‐ or 1.0‐mL sampling volume. Seven of 36 PCs with a 2.0‐mL sampling volume and 10 of 36 PCs with a 1.0‐mL sampling volume contained>10(6) CFU per mL of S. epidermidis at the time of first detection. CONCLUSION: Data from this preliminary evaluation suggest that sampling times of 24 hours or more would be necessary to provide confidence in detection of E. coli or S. epidermidis in PCs using this culture method.Keywords
This publication has 26 references indexed in Scilit:
- Rapid identification of bacterially contaminated platelets using reagent strips: glucose and pH analysis as markers of bacterial metabolismTransfusion, 1997
- Comparison of bacteria growth in single and pooled platelet concentrates after deliberate inoculation and storageTransfusion, 1995
- A prospective study of symptomatic bacteremia following platelet transfusion and of its managementTransfusion, 1994
- BACTERIAL CONTAMINATION OF CELLULAR BLOOD COMPONENTSVox Sanguinis, 1994
- A prospective microbiologic surveillance program to detect and prevent the transfusion of bacterially contaminated plateletsTransfusion, 1993
- Strategies for the avoidance of bacterial contamination of blood componentsTransfusion, 1993
- Staphylococcus epidermidis bacteremia from transfusion of contaminated platelets: application of bacterial DNA analysisTransfusion, 1992
- Prestorage removal of Yersinia enterocolitica from red cells with white cell‐reduction filtersTransfusion, 1992
- Yersinia Enterocolitica Serotype O:3 Sepsis after Blood TransfusionNew England Journal of Medicine, 1989
- Evaluation of the buffy-coat smear for rapid detection of bacteremia1981