Assessment of on-screen measurements, magnification, and calibration in digital radiography
Authors:
- Robyn L. ReadRRRobyn L. Read
- Colleen G. DuncanCDColleen G. Duncan
- Aaron D. WallaceAWAaron D. Wallace
- James A. PerryJPJames A. Perry
- Felix M. DuerrFDFelix M. Duerr
- 15 September 2012
- journal article
- Published by American Veterinary Medical Association (AVMA) in Journal of the American Veterinary Medical Association
- Vol. 241 (6) , 782-787
Abstract
Objective—To compare calibration methods for digital radiography in terms of measurement accuracy and interobserver variability. Design—Prospective study. Sample—Digital radiographic images of a 155-mm-long Steinmann pin. Procedures—Measurement of pin length on digital radiographs was determined with a 25.4-mm-diameter calibration ball and commercially available software program via 3 calibration methods (ie, no calibration, autocalibration, and manual calibration). Digital radiographs of the calibration ball and pin were obtained with each placed at various vertical heights from the table (7 heights) and horizontal distances from the center of the beam (4 distances). Measurements of pin length on digital radiographs were made by 4 observers who were blinded to the orientation of the calibration ball and pin. Results—Pin lengths obtained by each calibration method were significantly different from each other and from the true value. Manual calibration was the most accurate. There was no significant interobserver variability in measurements. There was no significant change in measurements when the calibration ball was moved horizontally, but pin length measurements changed significantly when the ball was moved vertically (away from the table) with an approximate magnification error of 1% per centimeter of distance between the calibration ball and pin. Conclusions and Clinical Relevance—For digital radiography, manual calibration is recommended to achieve the most accurate measurements. Ideally, the calibration ball should be placed at the same vertical height as the object to be measured; however, if this cannot be achieved, the magnification error can be expected to be approximately 1% per centimeter of distance.Keywords
This publication has 11 references indexed in Scilit:
- Determination of digitised radiograph magnification factors for pre-operative templating in hip prosthesis surgerySkeletal Radiology, 2009
- The reliability and accuracy of digital templating in total knee replacementThe Journal of Bone and Joint Surgery. British volume, 2009
- Evaluation of the Accuracy and Use of X-Ray Markers in Digital Templating for Total Hip ArthroplastyThe Journal of Arthroplasty, 2009
- Disc calibration for digital templating in hip replacementThe Journal of Bone and Joint Surgery. British volume, 2008
- ACQUISITION HARDWARE FOR DIGITAL IMAGINGVeterinary Radiology & Ultrasound, 2008
- VERTEBRAL HEART SIZE IN RETIRED RACING GREYHOUNDSVeterinary Radiology & Ultrasound, 2007
- Accurate scaling of digital radiographs of the pelvisThe Journal of Bone and Joint Surgery. British volume, 2006
- Assessment of reproducibility and accuracy in templating hybrid total hip arthroplasty using digital radiographsThe Journal of Bone and Joint Surgery. British volume, 2006
- Digital correction of magnification in pelvic x rays for preoperative planning of hip joint replacements: Theoretical development and clinical results of a new protocolMedical Physics, 2005
- COMPARISON OF ULTRASONOGRAPHIC AND RADIOGRAPHIC MEASUREMENTS OF CANINE PROSTATE DIMENSIONSVeterinary Radiology & Ultrasound, 1999