Determination of Cadmium in Water Samples by Automated Flow-Batch Cloud Point Extraction (CPE) Hyphenated to High-Resolution Continuum Source Flame Atomic Spectrometry (HR-CS FAAS)
Authors:
- 20 August 2021
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 55 (5) , 741-754
Abstract
A computer-controlled multicommuted flow-batch system is reported that allows for the online cloud point extraction of metal ions followed by high-resolution continuum source flame atomic absorption spectrometric (HR-CS GFAAS) detection. This approach combines reduced sample and reagent consumption and a short analysis time with the increased precision and sensitivity characteristic of flow-batch analysis and modern extraction techniques. As proof of concept, cadmium in aqueous solutions was determined by the formation of a hydrophobic complex [Cd(PAN)2]. Parameters related to complex formation (pH, pyridyl-azo-naphthol-PAN concentration, complexation time), cloud point extraction (Triton X-114 concentration, electrolyte composition, rich phase trapping), and the multicommuted flow-batch system (flow rates, timing control) were optimized. A small column packed with hydrophobic cotton was used as the interface between the extraction and detection stages to stack the cadmium-containing rich phase in a short plug. Optimized parameters led to a 5.0 to 75.0 µg L−1 linear range, 1.3 µg L−1 limit of detection, 4.3 µg L−1 limit of quantification and 1.8% coefficient of variation (n = 12). Accuracy was evaluated by analyzing a certified reference material EP-H (drinking water) and by recovery tests employing tap water, groundwater, mineral water, and reuse water samples.Keywords
Funding Information
- FAPESP (2011/19730-3, 2016/23878-0, 2015/11.523-0)
- CAPES
- financial support and Multiuser Central Facilities
- CNPq (308586/2018-2)
This publication has 41 references indexed in Scilit:
- Flow-batch analysisTrAC Trends in Analytical Chemistry, 2012
- Development of a cloud point extraction and preconcentration method for chromium(III) and total chromium prior to flame atomic absorption spectrometryJournal of Analytical Chemistry, 2012
- A preconcentration procedure for the determination of cadmium in biological material after on-line cloud point extractionEnvironmental Monitoring and Assessment, 2011
- Preconcentration and determination of trace metals by flow injection micelle-mediated extraction using flame atomic absorption spectrometryTalanta, 2009
- Mixing chambers in flow analysis: A reviewJournal of Analytical Chemistry, 2009
- Biological half-life of cadmium in the urine of inhabitants after cessation of cadmium exposureBiomarkers, 2009
- Simultaneous flow injection preconcentration of lead and cadmium using cloud point extraction and determination by atomic absorption spectrometryJournal of Hazardous Materials, 2009
- Multi-commutation in flow analysis: Recent developments and applicationsAnalytica Chimica Acta, 2008
- On-line metals preconcentration and simultaneous determination using cloud point extraction and inductively coupled plasma optical emission spectrometry in water samplesAnalytica Chimica Acta, 2008
- Review of procedures involving separation and preconcentration for the determination of cadmium using spectrometric techniquesJournal of Hazardous Materials, 2007