Detection of Sleep Disordered Breathing and Its Central/Obstructive Character Using Nasal Cannula and Finger Pulse Oximeter
Authors:
- Dirk SommermeyerDSDirk Sommermeyer
- Ding ZouDZDing Zou
- Ludger GroteLGLudger Grote
- Jan HednerJHJan Hedner
- 15 October 2012
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Sleep Medicine
- Vol. 8 (5) , 527-533
Abstract
To assess the accuracy of novel algorithms using an oximeter-based finger plethysmographic signal in combination with a nasal cannula for the detection and differentiation of central and obstructive apneas. The validity of single pulse oximetry to detect respiratory disturbance events was also studied. Patients recruited from four sleep laboratories underwent an ambulatory overnight cardiorespiratory polygraphy recording. The nasal flow and photoplethysmographic signals of the recording were analyzed by automated algorithms. The apnea hypopnea index (AHI(auto)) was calculated using both signals, and a respiratory disturbance index (RDI(auto)) was calculated from photoplethysmography alone. Apnea events were classified into obstructive and central types using the oximeter derived pulse wave signal and compared with manual scoring. Sixty-six subjects (42 males, age 54 ± 14 yrs, body mass index 28.5 ± 5.9 kg/m(2)) were included in the analysis. AHI(manual) (19.4 ± 18.5 events/h) correlated highly significantly with AHI(auto) (19.9 ± 16.5 events/h) and RDI(auto) (20.4 ± 17.2 events/h); the correlation coefficients were r = 0.94 and 0.95, respectively (p < 0.001) with a mean difference of -0.5 ± 6.6 and -1.0 ± 6.1 events/h. The automatic analysis of AHI(auto) and RDI(auto) detected sleep apnea (cutoff AHI(manual) ≥ 15 events/h) with a sensitivity/specificity of 0.90/0.97 and 0.86/0.94, respectively. The automated obstructive/central apnea indices correlated closely with manually scoring (r = 0.87 and 0.95, p < 0.001) with mean difference of -4.3 ± 7.9 and 0.3 ± 1.5 events/h, respectively. Automatic analysis based on routine pulse oximetry alone may be used to detect sleep disordered breathing with accuracy. In addition, the combination of photoplethysmographic signals with a nasal flow signal provides an accurate distinction between obstructive and central apneic events during sleep.Keywords
This publication has 33 references indexed in Scilit:
- The European Sleep Apnoea Database (ESADA): report from 22 European sleep laboratoriesEuropean Respiratory Journal, 2011
- Portable monitoring in sleep apnoea: the way forward?European Respiratory Journal, 2011
- Association of Incident Cardiovascular Disease With Progression of Sleep-Disordered BreathingCirculation, 2011
- Prognostic Impact of Sleep Disordered Breathing and Its Treatment in Heart Failure: An Observational StudyEuropean Journal of Heart Failure, 2011
- Pulse Wave Amplitude Drops during Sleep are Reliable Surrogate Markers of Changes in Cortical ActivitySLEEP, 2010
- Comparison of the automatic analysis versus the manual scoring from ApneaLink™ device for the diagnosis of obstructive sleep apnoea syndromeSleep and Breathing, 2010
- Noninvasive Cardiovascular Markers of Acoustically Induced Arousal from Non–rapid-eye-movement SleepSLEEP, 2002
- Autonomic Arousal Index: an Automated Detection Based on Peripheral Arterial TonometrySLEEP, 2002
- Epidemiology of Obstructive Sleep Apnea: A Population Health PerspectiveAmerican Journal of Respiratory and Critical Care Medicine, 2002
- Overnight Pulse Oximetry for Sleep-Disordered Breathing in AdultsChest, 2001