Comparison of four tonometers in the measurement of intraocular pressure in healthy horses
- 16 January 2023
- journal article
- research article
- Published by Wiley in Equine Veterinary Journal
- Vol. 55 (6) , 1104-1111
Abstract
Background: Measurement of the intraocular pressure (IOP) is a useful diagnostic tool in equine ophthalmology. Handheld tonometers, such as Tonovet and Tonovet Plus (rebound), Tono‐Pen AVIA Vet (applanation), and Kowa HA‐2 (applanation using the Goldmann methodology) are used to obtain IOP measurements in veterinary medicine.Objectives: To compare and evaluate the accuracy of four handheld tonometers in measuring IOP using different methodologies in healthy horses.Study design: In vivo experiment and cross‐sectional survey of healthy horses.Methods: Intraocular pressure was measured in 72 eyes of 36 horses. An in vivo study was conducted on sedated horses to compare the real IOP values obtained using manometry versus those obtained using tonometry, and a field study was conducted on unsedated healthy horses with normal eyes to measure the IOP values using different tonometers.Results: In the in vivo study, the mean IOP values using ocular manometry was 24.9 ± 4.0 mmHg (range, 20.0–30.0 mmHg). The mean IOP values using tonometry were: Tonovet, 25.7 ± 5.8 mmHg (range 19.5–33.0 mmHg); Tonovet Plus, 24.8 ± 7.1 mmHg (range 13.2–33.2 mmHg); Tono Pen AVIA Vet, 19.2 ± 4.7 mmHg (range 13.1–26.5 mmHg); and Kowa Ha‐2, 24.1 ± 1.2 mmHg (range 22.8–25.8 mmHg). In the field study, the IOP values were: Tonovet, 30.7 ± 5.6 mmHg (range 21.7–38.0 mmHg); Tonovet Plus, 29.6 ± 6.7 mmHg (range 16.2–38.6 mmHg); Tono‐Pen AVIA Vet, 27.3 ± 5.8 mmHg (range 14.6–37.1 mmHg); and Kowa HA‐2, 23.4 ± 2.2 mmHg (range 20.2–28.7 mmHg).Main limitations: This study included only healthy horses and a limited number of animals in the in vivo study.Conclusions: There was a strong correlation between the IOP values and manometry for all tonometers. IOP should be estimated using the same tonometer over time, and the bias of the tonometer used, such as overestimation (rebound tonometer) and underestimation (applanation tonometer), should be acknowledged. A normal reference value for each tonometer should be established in horses.Keywords
This publication has 25 references indexed in Scilit:
- Intraocular pressure measurements with the Tono-Pen XL® and Perkins® applanation tonometers in horses and cattleCiência Rural, 2013
- Evaluation of the Perkins handheld applanation tonometer in horses and cattleJournal of Veterinary Science, 2011
- Accuracy and reproducibility of the TonoVet®rebound tonometer in birds of preyVeterinary Ophthalmology, 2010
- Circadian Intraocular Pressure Rhythms in Athletic Horses under Different Lighting RegimeChronobiology International, 2009
- Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thicknessBritish Journal of Ophthalmology, 2006
- Effect of head position on intraocular pressure in horsesAmerican Journal of Veterinary Research, 2006
- Comparison of the rebound tonometer (ICare®) to the applanation tonometer (Tonopen XL®) in normotensive dogsVeterinary Ophthalmology, 2006
- Steroid-Induced Ocular Hypertension in Normal CattleArchives of Ophthalmology (1950), 2004
- Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anestheticGraefe's Archive for Clinical and Experimental Ophthalmology, 2003
- Efeito de drogas utilizadas no tratamento de hipertensão arterial sistêmica sobre a pressão intra-ocular: estudo experimental no cãoArquivos Brasileiros de Oftalmologia, 2002