Phantom studies in osteoporosis
Authors:
- 1 May 1993
- journal article
- occasional survey
- Published by Springer Nature in European Journal of Nuclear Medicine and Molecular Imaging
- Vol. 20 (5) , 434-439
Abstract
Differences in measurement results of bone densitometry are an obvious disadvantage of this method. The differences are mainly due to the calibration procedures for bone densitometry systems employed by the manufacturers, the software algorithms for defining the region of interest or edge detection, and the physiological inhomogeneity of body composition. Whereas intea-unit variation of reproducibility is acceptable, inter-unit variation may reach up to 20%. This paper discusses the problems of designing measurement phantoms and underlines the need for standardisation of phantoms for calibration, cross-calibration, and quality control in bone densitometry. A general phantom used for cross-calibration should handle all parameters influencing measurement of bone minerals to yieldet dynamic reference values. One has to note that densitometry systems do not measure the absolute bone mineral content but a model-related equivalent of the calibration material.Keywords
This publication has 12 references indexed in Scilit:
- Phantoms for evaluating the performance characteristics of bone densitometersBritish Journal of Radiology, 1992
- Photon absorption measurements of bone density in the presence of scattered radiationPhysics in Medicine & Biology, 1992
- Errors due to non-uniform distribution of fat in dual X-ray absorptiometry of the lumbar spineBritish Journal of Radiology, 1992
- A phantom for standarization and quality control in spinal bone mineral measurements by QCT and DXA: Design considerations and specificationsMedical Physics, 1992
- Dual energy x-ray absorptiometry: The effects of beam hardening on bone density measurementsMedical Physics, 1992
- A comparison of two dual-energy X-ray absorptiometry systems for spinal bone mineral measurementCalcified Tissue International, 1992
- Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometersBritish Journal of Radiology, 1991
- Intersite comparison of the Hologic QDR-1000 dual energy X-ray bone densitometerBritish Journal of Radiology, 1991
- Vergleichende Messung des Knochenmineralgehaltes mit DPA und DPX - Erste klinische ErfahrungenRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1990
- Dual Photon Absorptiometky in Lumbar Vertebrae: I. Theory and MethodActa Radiologica: Therapy, Physics, Biology, 1974