Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture
Authors:
Open Access
- 1 June 2018
- journal article
- research article
- Published by Springer Nature in European Journal of Plastic Surgery
- Vol. 41 (5) , 563-570
Abstract
Background The Poly Implant Prothèse (PIP) implants were withdrawn from the market in 2010 due to the use of low-grade silicone, causing a high risk for implant rupture. The aim of this study was to investigate the implant dynamics of PIP breast implants, as well as to determine the rate and predictors of implant gel bleeding, rupture, and capsular contracture in PIP implants. Methods Eighty women with a total of 152 PIP implants who underwent a reoperation in 2012 were enrolled in this study. Physical investigation included assessing the Baker score and demographics were retrospectively traced in medical records. The pre- and post-operative volumes of the implants were calculated and their state was determined intraoperatively by the surgeon. Results The implants were removed after a mean implant duration of 11 ± 2.1 years. Gel bleed and implant rupture occurred in respectively 42 and 25% of the implants. Intact implants had post-operative volume increase as well as decrease. There was a correlation between gel bleeding and more post-operative implant volume increase (P ≤ 0.05). Capsular contracture had a protective effect against post-operative implant volume increase (P ≤ 0.05), while a post-operative implant volume increase provided a protective influence in developing capsular contracture (P ≤ 0.05). Additionally, implant rupture led to a higher risk of capsular contracture (P ≤ 0.05). Conclusions We managed to illustrate that PIP implant shells were too permeable and that there is a correlation between gel bleeding and the increase of the post-operative implant volume. Implant rupture led to a higher risk for developing capsular contracture. Level of evidence: Level III, risk / prognostic study.Keywords
Funding Information
- VU University Medical Center
This publication has 32 references indexed in Scilit:
- Poly Implant Prothèse (PIP) breast implants: Our experienceThe Surgeon, 2013
- The Role of Silicone Granulomas in Recurrent Capsular ContracturePlastic and Reconstructive Surgery, 2013
- Chemical and physicochemical properties of the high cohesive silicone gel from Poly Implant Prothèse (PIP) breast prostheses after explantation: A preliminary, comparative analytical investigationJournal of Pharmaceutical and Biomedical Analysis, 2013
- Bilan à 18 mois du retrait du marché des prothèses Poly Implant Prothèse. Expérience d’un centre anticancéreuxAnnales de Chirurgie Plastique Esthétique, 2012
- Bacterial Colonization Is of Major Relevance for High-Grade Capsular Contracture After Augmentation MammaplastyAnnals of Plastic Surgery, 2007
- Cellular and molecular composition of fibrous capsules formed around silicone breast implants with special focus on local immune reactionsJournal of Autoimmunity, 2004
- Epidemiological Investigation of Local Complications After Cosmetic Breast Implant Surgery in DenmarkAnnals of Plastic Surgery, 2002
- Analysis of Risk Factors Associated with Rupture of Silicone Gel Breast ImplantsPlastic and Reconstructive Surgery, 1999
- Analysis of Risk Factors Associated with Rupture of Silicone Gel Breast ImplantsPlastic and Reconstructive Surgery, 1999
- Classification of Capsular Contracture after Prosthetic Breast ReconstructionPlastic and Reconstructive Surgery, 1995