Impact of four kVp combinations available in a dual‐source CT on the spectral performance of abdominal imaging: A task‐based image quality assessment on phantom data
Authors:
Open Access
- 26 July 2021
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 22 (8) , 243-254
Abstract
Purpose To compare the spectral performance of four combinations of kVp available in a third generation dual-source CT (DSCT) on abdominal imaging. Methods An image-quality phantom was scanned with a DSCT using four kVp pairs (tube “A” voltage/tube “B” voltage): 100/Sn150 kVp, 90/Sn150 kVp, 80/Sn150 kVp, and 70/Sn150 kVp, classic parameters and dose level for abdomen examination (CTDIvol: 11 mGy). The noise power spectrum (NPS) and the task-based transfer function (TTF) of two inserts were computed on virtual monochromatic images (VMIs) at 40/50/60/70 keV and for mixed, low-, and high-kVp images. Detectability index (d’) was computed on VMIs and mixed images to model the detection task of liver metastasis (LM) and hepatocellular carcinoma (HCC). Iodine quantification accuracy was assessed using the Root Mean Square Deviation (RMSDiodine) and the iodine bias (IB). Results Noise magnitude decreased by −55%± 0% between 40 and 70 keV for all kVp pairs. Compared to 70/Sn150 kVp, noise magnitude was increased by 9% ± 0% with 80/Sn150 kVp, by 16% ± 1% with 90/Sn150 kVp and by 24%± 1% with 100/Sn150 kVp. The average NPS spatial frequency (fav) shifted toward higher frequencies as energy level increased for all kVp pairs. Lowest fav values were found for 70/Sn150 kVp and highest for 100/Sn150 kVp. The value of TTF at 50% (f50) shifted toward lower frequencies with increasing energy level. The highest f50 values occurred for 100/Sn150 kVp and the lowest for 80/Sn150 kVp. For both lesions, d’ was highest for 70/Sn150 kVp and lowest for 100/Sn150 kVp. Compared to 70/Sn150 kVp, d’ decreased by −6% ± 3% with 80/Sn150 kVp, by −11% ± 2% with 90/Sn150 kVp and by −13%± 2% with 100/Sn150 kVp. For all acquisitions, the RSMDiodine and IB were the lowest for 100/Sn150 kVp (0.29 ± 0.10 mg/ml and 0.88 ± 0.30 mg/ml, respectively) and increased when the tube “A” voltage decreased (2.34 ± 0.29 mg/ml for 70/Sn150 kVp and 7.42 ± 0.51 mg/ml respectively). Conclusion 70/Sn150 kVp presented the lowest image noise and highest detectability in VMIs of two small focal liver lesions. 100/Sn150 kVp presented the lowest image noise on mixed images and highest accuracy of iodine quantification in iodine images.Keywords
This publication has 33 references indexed in Scilit:
- Intermanufacturer Comparison of Dual-Energy CT Iodine Quantification and Monochromatic Attenuation: A Phantom StudyRadiology, 2018
- Low kV versus dual-energy virtual monoenergetic CT imaging for proven liver lesions: what are the advantages and trade-offs in conspicuity and image quality? A pilot studyAbdominal Radiology, 2017
- Selection of optimal tube potential settings for dual-energy CT virtual mono-energetic imaging of iodine in the abdomenAbdominal Radiology, 2017
- Principles and Clinical Application of Dual-energy Computed Tomography in the Evaluation of Cerebrovascular DiseaseJournal of Clinical Imaging Science, 2016
- Applications of dual energy computed tomography in abdominal imagingDiagnostic and Interventional Imaging, 2016
- Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical ApplicationsRadiology, 2015
- The Importance of Spectral SeparationInvestigative Radiology, 2015
- State of the Art: Dual-Energy CT of the AbdomenRadiology, 2014
- Dual-Energy CT Applications in the AbdomenAmerican Journal of Roentgenology, 2012
- Dual-Source Dual-Energy CT With Additional Tin Filtration: Dose and Image Quality Evaluation in Phantoms and In VivoAmerican Journal of Roentgenology, 2010