Validation rules for blood smear revision after automated hematological testing using Mindray CAL-8000
Authors:
- 6 October 2016
- journal article
- research article
- Published by Wiley in Journal of Clinical Laboratory Analysis
- Vol. 31 (4) , e22067
Abstract
This article was aimed to test the use of validation rules for blood smear review after automated hematological testing using Mindray CAL-8000 (two hematological analyzers and one autoslider). This study was based on 1013 peripheral blood samples (PB) referred for routine hematological testing. Results of testing on CAL-8000 were analyzed using both locally derived and International Consensus Group for Hematology (ICGH) validation rules, and then compared with data obtained by optical microscopy (OM). A workflow analysis was also completed. The overall agreement with locally derived and ICGH criteria was 91% and 85%, but a higher sensitivity was observed for locally derived criteria (0.97 vs 0.95). The percentage of false negative and false positive samples was 2.1% and 7.1% using ICGH criteria, and was 1.4% and 14% using locally defined rules. The throughput of CAL-8000 system was 208 samples/h, with a percentage of OM analysis comprised between 14% and 17%, and sensitivity of 0.97. As regards personnel activity, we estimated 0.8 full-time equivalent (FTE) of technical staff and 0.7 FTE of personnel for clinical validation of data and blood smear review. These results show that customization of validation rules is necessary for enhancing the quality of hematological testing and optimizing workflow.Keywords
This publication has 18 references indexed in Scilit:
- Comparison of Automated Differential Blood Cell Counts From Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in Normal and Pathologic SamplesAmerican Journal of Clinical Pathology, 2013
- Red blood cell morphologyInternational Journal of Laboratory Hematology, 2013
- Validation and Optimization of Criteria for Manual Smear Review Following Automated Blood Cell Analysis in a Large University HospitalArchives of Pathology & Laboratory Medicine, 2013
- Purpose and Criteria for Blood Smear Scan, Blood Smear Examination, and Blood Smear ReviewAnnals of Laboratory Medicine, 2013
- EDTA-dependent pseudothrombocytopenia: further insights and recommendations for prevention of a clinically threatening artifactClinical Chemistry and Laboratory Medicine (CCLM), 2012
- Evaluation of the Beckman Coulter UniCel DxH 800, Beckman Coulter LH 780, and Abbott Diagnostics Cell-Dyn Sapphire Hematology Analyzers on Adult Specimens in a Tertiary Care HospitalAmerican Journal of Clinical Pathology, 2011
- Optimization of laboratory workflow in clinical hematology laboratory with reduced manual slide review: comparison between Sysmex XE-2100 and ABX Pentra DX120International Journal of Laboratory Hematology, 2011
- Automated Blood Cell CountsAmerican Journal of Clinical Pathology, 2008
- The International Consensus Group for Hematology Review: Suggested Criteria for Action Following Automated CBC and WBC Differential AnalysisLaboratory Hematology, 2005
- Quality specification in haematology: the automated blood cell countClinica Chimica Acta, 2004