Steady state of gradient echo sequences with radiofrequency phase cycling: Analytical solution, contrast enhancement with partial spoiling
Author:
Open Access
- 9 December 2005
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 55 (1) , 98-107
Abstract
Spoiled gradient echo sequences can only reach a homogeneous steady state if sufficiently strong crusher gradients are used in combination with RF phase cycling (RF spoiling). However, the signal depends quite sensitively on the chosen phase increment ϕ and—lacking analytical solutions—numerical simulations must be used to study the transient and steady‐state magnetization. For the steady state an exact analytical solution is derived, which holds for arbitrary sequence and tissue parameters. Besides a considerably improved computation performance, the analytical approach enables a better understanding of the complicated dependence on ϕ. For short repetition times (TR) the regime of small ϕ turns out to be particularly interesting: It is shown that the typical ϕc, where RF spoiling starts to become effective, is essentially inversely proportional to T2. This tissue dependence implies that contrasts can be considerably larger with partial spoiling (ϕ ≈ ϕc) than with conventional RF spoiling (ϕ ≫ ϕc). As an example, the uptake of contrast agents in tissues is investigated. For typical parameters a considerably improved contrast enhancement can be obtained, both theoretically and experimentally. Magn Reson Med, 2006.Keywords
This publication has 11 references indexed in Scilit:
- Off‐resonance effects in the transient response of SSFP sequencesMagnetic Resonance in Medicine, 2004
- Steady‐state preparation for spoiled gradient echo imagingMagnetic Resonance in Medicine, 2001
- Spoiling of transverse magnetization in gradient‐echo (GRE) imaging during the approach to steady stateMagnetic Resonance in Medicine, 1996
- Spoiling of transverse magnetization in steady‐state sequencesMagnetic Resonance in Medicine, 1991
- Echoes—how to generate, recognize, use or avoid them in MR‐imaging sequences. Part I: Fundamental and not so fundamental properties of spin echoesConcepts in Magnetic Resonance, 1991
- Elimination of transverse coherences in FLASH MRIMagnetic Resonance in Medicine, 1988
- Artifacts due to residual magnetization in three‐dimensional magnetic resonance imagingMedical Physics, 1988
- Steady-State Magnetizations in Rapid NMR Imaging Using Small Flip Angles and Short Repetition IntervalsIEEE Transactions on Medical Imaging, 1987
- Diffusion and field-gradient effects in NMR Fourier spectroscopyThe Journal of Chemical Physics, 1974
- Effects of Diffusion in Nuclear Magnetic Resonance Spin-Echo ExperimentsThe Journal of Chemical Physics, 1961