A comparative study for robot assisted vitreoretinal surgery: Micron vs. the Steady-Hand Robot
Open Access
- 1 May 2013
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 10504729,p. 4832-4837
Abstract
In vitreoretinal surgery, application of excessive forces and unintentional motion due to hand-tremor can easily result in serious complications. Robotic assistance when combined with tool-to-tissue force sensing capabilities has significant potential to improve such practice. In this paper, we evaluate the membrane peeling performance of a single user for two distinct robotic systems with integrated force sensing capabilities: Micron and the Steady-Hand Robot. We show that these systems provide promising performance improvement with similar impact on peeling forces and comparable tremor cancellation trends.Keywords
This publication has 16 references indexed in Scilit:
- Force sensing micro-forceps for robot assisted retinal surgery2012
- Force sensing micro-forceps with integrated fiber Bragg grating for vitreoretinal surgery2012
- Robot-Assisted Vitreoretinal SurgeryOphthalmology, 2009
- Sub-retinal hemorrhage during internal limiting membrane peeling for a macular holeGraefe's Archive for Clinical and Experimental Ophthalmology, 2003
- Evaluation of a Telerobotic System to Assist Surgeons in MicrosurgeryComputer Aided Surgery, 1999
- Characterization of Epiretinal Membranes Using Optical Coherence TomographyOphthalmology, 1996
- Distribution of Iatrogenic Retinal Breaks in Macular Hole SurgeryOphthalmology, 1995
- Ophthalmic microsurgical robot and associated virtual environmentComputers in Biology and Medicine, 1995
- A Teleoperated Microsurgical Robot and Associated Virtual Environment for Eye SurgeryPRESENCE: Virtual and Augmented Reality, 1993
- Iatrogenic Retinal Breaks Complicating Pars Plana VitrectomyOphthalmology, 1990