Performance of Mycobacterium Growth Indicator Tube BACTEC 960 with Lowenstein–Jensen method for diagnosis of Mycobacterium tuberculosis at Ethiopian National Tuberculosis Reference Laboratory, Addis Ababa, Ethiopia
Authors:
- Getu DiribaGDGetu Diriba
- Abebaw KebedeAKAbebaw Kebede
- Zelalem YaregalZYZelalem Yaregal
- Muluwork GetahunMGMuluwork Getahun
- Mengistu TadesseMTMengistu Tadesse
- Abyot MeazaAMAbyot Meaza
- Zekarias DagneZDZekarias Dagne
- Shewki MogaSMShewki Moga
- Jibril DileboJDJibril Dilebo
- Kebebe GudenaKGKebebe Gudena
- Mulu HassenMHMulu Hassen
- Kassu DestaKDKassu Desta
Open Access
- 10 May 2017
- journal article
- research article
- Published by Springer Nature in BMC Research Notes
- Vol. 10 (1) , 1-6
Abstract
Bacteriological confirmed active case detection remains the corner stone for diagnosing tuberculosis. Non-radiometric liquid culture system Mycobacterium Growth Indicator Tube with automated interface had been recommended by expert groups in addition to conventional solid culture media such as Lowenstein–Jensen. However in high burden resource limited countries advanced non-radiometric based tuberculosis diagnostic methods such as MGIT 960 is limited. Therefore we have evaluated the performance of MGIT 960 system compared to LJ for recovery of Mycobacterium complex (MTBC) from clinical specimens. A cross sectional study was conducted from a total of 908 samples between January 1st, 2013 to December 31st, 2014. Clinical specimens were processed following standard procedures and the final suspension was inoculated to MGIT tubes and LJ slant. Identification and confirmation of MTBC was done by ZN staining and SD Bioline test. Data was analyzed by SPSS version 20. The sensitivity, specificity, recovery rate and the average turnaround time to recover the organism was computed. From a total of 908 clinical specimens processed using both LJ and BACTEC MGIT liquid culture methods the recovery rate for LJ and MGIT, for smear positive samples was 66.7% (74/111) and 87.4% (97/ 111) respectively while for smear negative samples was 13.4% (108/797) and 17.4% (139/797) for LJ and MGIT methods respectively. The overall recovery rate for MGIT is significantly higher than LJ methods [26% (236/908; vs. 20%, 182/908, P = 0.002)]. The average turnaround time for smear positive samples was 16 and 31 days for MGIT and LJ respectively. Turnaround time for smear negative samples was 20 and 36 days for MGIT and LJ respectively. The overall agreement between MGIT and LJ was fairly good with Kappa value of 0.59 (P < 0.001). In the present study the contamination rate for MGIT is higher than the LJ methods, 15 and 9.3% respectively. The BACTEC MGIT liquid culture system has better MTBC recovery rate with shorter turnaround time for both smear positive and negative clinical specimens compared to Conventional LJ method. However, efforts should be made in order to reduce the high contamination rate in BACTEC MGIT system and to lesser extent to LJ methods.Keywords
This publication has 26 references indexed in Scilit:
- Comparison of the performance of TK system with LJ and MGIT methods in the diagnosis of tuberculosis.2014
- Comparison of Mycobacterium tuberculosis drug susceptibility using solid and liquid culture in NigeriaBMC Research Notes, 2013
- Iron Deposition and Ferritin Heavy Chain (Fth) Localization in Rodent TeethBMC Research Notes, 2013
- The manual MGIT system for the detection of M. tuberculosis in respiratory specimens: an experience in the University Malaya Medical Centre.2009
- Improvement in Mycobacterial Yield and Reduced Time to Detection in Pediatric Samples by Use of a Nutrient Broth Growth SupplementJournal of Clinical Microbiology, 2009
- Molecular diagnostics in tuberculosisEuropean Journal of Clinical Microbiology & Infectious Diseases, 2005
- Comparison of the conventional diagnostic modalities, bactec culture and polymerase chain reaction test for diagnosis of tuberculosisIndian Journal of Medical Microbiology, 2005
- Comparison of the Automated Mycobacteria Growth Indicator Tube System (BACTEC 960/MGIT) With Löwenstein-Jensen Medium for Recovery of Mycobacteria From Clinical SpecimensAmerican Journal of Clinical Pathology, 2002
- Evaluation of BACTEC MGIT 960 and BACTEC 460TB Systems for Recovery of Mycobacteria from Clinical Specimens of a University Hospital with Low Incidence of TuberculosisJournal of Clinical Microbiology, 2001
- Comparison of MB/BacT ALERT 3D System with Radiometric BACTEC System and Löwenstein-Jensen Medium for Recovery and Identification of Mycobacteria from Clinical Specimens: a Multicenter StudyJournal of Clinical Microbiology, 2001