Evaluation of a photon and an electron beam of a 6-MV linear accelerator
Authors:
- Prabakar ModurPMPrabakar Modur
- Ram BasavatiaRBRam Basavatia
- 1 July 1988
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 15 (4) , 525-529
Abstract
The first Mitsubishi medical linear accelerator in the United States was commissioned in April 1985. This unit EXL-8 (marketed by Mitsubishi International Corporation) produces 8-MeV electron beams in addition to 6-MV X rays. It is a 100-cm source-axis distance isocentric machine. Acceptance testing and performance evaluation of this accelerator were completed. Our measurement included beam characteristics and dosimetry parameters for both modalities. Central axis % depth dose (% DD), tissue-maximum ratio, field size output factors, wedge factors, etc., for this Linac 6-MV beam are reported. Characteristics of the 8-MeV electron beam, namely % DD data, isodose curves, and cone ratios for various electron applicators are presented.Keywords
This publication has 5 references indexed in Scilit:
- Dosimetry of a dual photon energy linear acceleratorMedical Physics, 1987
- Electron beam central axis depth dose measurementsMedical Physics, 1985
- Characteristics of an MM 22 medical microtron 6-MV photon beamMedical Physics, 1984
- Buildup region and skin‐dose measurements for the Therac 6 Linear Accelerator for radiation therapyMedical Physics, 1981
- Effect of Air Space on Depth Dose in Electron Beam TherapyRadiology, 1978