Biomechanical study of 3D-printed titanium alloy pad for repairing glenoid bone defect
Authors:
Open Access
- 20 May 2024
- journal article
- research article
- Published by SAGE Publications in Journal of Orthopaedic Surgery
- Vol. 32 (2)
Abstract
Background: The purpose of this study was to investigate the effect of 3D-printed technology to repair glenoid bone defect on shoulder joint stability. Methods: The shoulder joints of 25 male cadavers were tested. The 3D-printed glenoid pad was designed and fabricated. The specimens were divided into 5 groups. Group A: no bone defect and the structure of the glenoid labrum and joint capsule was intact; Group B: Anterior inferior bone defect of the shoulder glenoid; Group C: a pad with a width of 2 mm was installed; Group D: a pad with a width of 4 mm was installed; Group E: a pad with a width of 6 mm was installed. This study measured the distance the humeral head moved forward at the time of glenohumeral dislocation and the maximum load required to dislocate the shoulder. Results: The shoulder joint stability and humerus displacement was significantly lower in groups B and C compared with group A ( p < .05). Compared with group A, the stability of the shoulder joint of group D was significantly improved ( p < .05). However, there was no significant difference in humerus displacement between groups D and A ( p > .05). In addition, compared with group A, shoulder joint stability was significantly increased and humerus displacement was significantly decreased in group E ( p < .05). Conclusion: The 3D-printed technology can be used to make the shoulder glenoid pad to perfectly restore the geometric shape of the shoulder glenoid articular surface. Moreover, the 3D-printed pad is 2 mm larger than the normal glenoid width to restore the initial stability of the shoulder joint.Keywords
Funding Information
- Xiamen Science and Technology Plan Project (3502Z20194052)
- Xiamen Key Specialty construction project of Traumatic Or-thopedics department (2015347)
- Xiamen Superior Sub-specialty construction project of Arthro-scopic minimally invasive Orthopedics department (2018296)
This publication has 24 references indexed in Scilit:
- Titanium and Titanium Alloys2019
- A biomechanical shoulder strain index based on stabilizing demand of shoulder jointErgonomics, 2018
- Glenoid morphology is associated with the development of instability arthropathyJournal of Shoulder and Elbow Surgery, 2018
- Proper benefit of a three dimensional pre-operative planning software for glenoid component positioning in total shoulder arthroplastyInternational Orthopaedics, 2018
- Glenoid labrum pathologyOrthopaedics & Traumatology: Surgery & Research, 2015
- Approach to the treatment of primary anterior shoulder dislocation: A reviewThe Physician and Sportsmedicine, 2015
- Anatomic study of the superior glenoid labrumClinical Anatomy, 2012
- Radiocapitellar stability: the effect of soft tissue integrity on bipolar versus monopolar radial head prosthesesJournal of Shoulder and Elbow Surgery, 2011
- In vitro biocompatibility evaluation of surface‐modified titanium alloysJournal of Biomedical Materials Research Part A, 2009
- Hydroxyapatite porous coating and the osteointegration of the total hip replacementArchives of orthopaedic and trauma surgery, 2006