Three-Dimensional Endovascular Navigation With Electromagnetic Tracking: Ex Vivo and In Vivo Accuracy
- 1 April 2011
- journal article
- research article
- Published by International Society of Endovascular Specialists in Journal of Endovascular Therapy
- Vol. 18 (2) , 230-240
Abstract
Purpose To evaluate the accuracy of a 3-dimensional (3D) navigation system using electromagnetically tracked tools to explore its potential in patients. Methods The 3D navigation accuracy was quantified on a phantom and in a porcine model using the same setup and vascular interventional suite. A box-shaped phantom with 16 markers was scanned in 5 different positions using computed tomography (CT). The 3D navigation system registered each CT volume in the magnetic field. A tracked needle was pointed at the physical markers, and the spatial distances between the tracked needle positions and the markers were calculated. Contrast-enhanced CT images were acquired from 6 swine. The 3D navigation system registered each CT volume in the magnetic field. An electromagnetically tracked guidewire and catheter were visualized in the 3D image and navigated to 4 specified targets. At each target, the spatial distance between the tracked guidewire tip position and the actual position, verified by a CT control, was calculated. Results The mean accuracy on the phantom was 1.28±0.53 mm, and 90% of the measured distances were ≤1.90 mm. The mean accuracy in swine was 4.18±1.76 mm, and 90% of the measured distances were ≤5.73 mm. Conclusion This 3D navigation system demonstrates good ex vivo accuracy and is sufficiently accurate in vivo to explore its potential for improved endovascular navigation.Keywords
This publication has 24 references indexed in Scilit:
- Electromagnetic Navigation for Thoracic Aortic Stent-graft Deployment: A Pilot Study in SwineJournal of Vascular and Interventional Radiology, 2010
- Intermediate results of a United States multicenter trial of fenestrated endograft repair for juxtarenal abdominal aortic aneurysmsJournal of Vascular Surgery, 2009
- Comparison of Suction Catheter versus Forceps Biopsy for Sampling of Solitary Pulmonary Nodules Guided by Electromagnetic Navigational BronchoscopyRespiration, 2009
- Branched devices for thoracoabdominal aneurysm repair: Early experienceJournal of Vascular Surgery, 2008
- Wholly endovascular repair of thoracoabdominal aneurysmBritish Journal of Surgery, 2008
- Endovascular Repair of Thoracoabdominal AneurysmsCirculation, 2008
- Endovascular treatment of thoracoabdominal aortic aneurysmsJournal of Vascular Surgery, 2008
- Side-Branched AAA Stent Graft Insertion Using Navigation Technology: A Phantom StudyEuropean Surgical Research, 2007
- Radiation exposure in stent-grafting of abdominal aortic aneurysmsBritish Journal of Radiology, 2005
- Three-dimensional navigated laparoscopic ultrasonographySurgical Endoscopy, 2001