The Thermodilution Technique for Measuring Resting Energy Expenditure Does Not Agree With Indirect Calorimetry for the Critically III Patient
Authors:
- Scott A. ShikoraSSScott A. Shikora
- Linda M. BurkeLBLinda M. Burke
- Jane E. Heetderks‐CoxJHJane E. Heetderks‐Cox
- Carl T. BergrenCBCarl T. Bergren
- Peter C. MuskatPMPeter C. Muskat
- 1 November 1998
- journal article
- research article
- Published by Wiley in Journal of Parenteral and Enteral Nutrition
- Vol. 22 (6) , 347-351
Abstract
Background: The complications associated with overfeeding critically ill patients are well documented. Indirect calorimetry is touted as the gold standard for measuring resting energy expenditure (REE). Unfortunately, the device is expensive, and many centers do not have this technology. The thermodilution technique for measuring cardiac output and calculating REE using the Fick equation has been reported to be an acceptable alternative. This study compared these techniques in a critically ill population. Methods: Forty consecutive patients with indwelling Swan—Ganz catheters in the surgical intensive care unit were prospectively studied while under the consultative care of the nutrition support service. REE was determined in all patients by both techniques within a 2‐hour period. An error of 5% (approximately ±100 kcal/d) between the two methods was deemed acceptable for clinical use. Results: Mean values for REE were 1928 ± 558 vs 1898 ± 518 kcal/d for the indirect calorimetry and thermodilution methods, respectively, and were not significantly different. However, there was great variation between the two techniques for the majority of patients such that REE determinations did not agree (t = 6.8; p <.0005). In 70% of the patients, REE determinations differed by ≥20% and in 10% of the patients by 50%. Additionally, the greater the difference between the two methods, the more the thermodilution method tended to overestimate REE. Conclusions: When compared with indirect calorimetry in a critically ill population, the thermodilution method demonstrated an intersubject variability that is unacceptable for clinical use. (Journal of Parenteral and Enteral Nutrition 22:347–351, 1998)Keywords
This publication has 18 references indexed in Scilit:
- Accuracy of 30‐Minute Indirect Calorimetry Studies in Predicting 24‐Hour Energy Expenditure in Mechanically Ventilated, Critically Ill PatientsJournal of Parenteral and Enteral Nutrition, 1997
- Clinical Comparison of Three Methods to Determine Resting Energy ExpenditureNutrition in Clinical Practice, 1994
- Energy expenditure and gas exchange measurements in postoperative patientsCritical Care Medicine, 1992
- Clinical application of the metabolic cart to the delivery of total parenteral nutritionCritical Care Medicine, 1990
- Peritoneal dialysis for acute renal failureCritical Care Medicine, 1990
- Validation of a compact system for measuring gas exchangeCritical Care Medicine, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Instrumentation for monitoring gas exchange and metabolic rate in critically ill patientsCritical Care Medicine, 1984
- A Simple and Accurate Indirect Calorimetry System for Assessment of Resting Energy ExpenditureJournal of Parenteral and Enteral Nutrition, 1984
- Glucose Requirements Following Burn InjuryAnnals of Surgery, 1979