Integration of immobilized trypsin bead beds for protein digestion within a microfluidic chip incorporating capillary electrophoresis separations and an electrospray mass spectrometry interface
- 25 July 2000
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 14 (15) , 1377-1383
Abstract
A microfluidic device is described in which an electrospray interface to a mass spectrometer is integrated with a capillary electrophoresis channel, an injector and a protein digestion bed on a monolithic substrate. A large channel, 800 µm wide, 150 µm deep and 15 mm long, was created to act as a reactor bed for trypsin immobilized on 40–60 µm diameter beads. Separation was performed in channels etched 10 µm deep, 30 µm wide and about 45 mm long, feeding into a capillary, attached to the chip with a low dead volume coupling, that was 30 mm in length, with a 50 µm i.d. and 180 µm o.d. Sample was pumped through the reactor bed at flow rates between 0.5 and 60 µL/min. The application of this device for rapid digestion, separation and identification of proteins is demonstrated for melittin, cytochrome c and bovine serum albumin (BSA). The rate and efficiency of digestion was related to the flow rate of the substrate solution through the reactor bed. A flow rate of 1 or 0.5 µL/min was found adequate for complete consumption of cytochrome c or BSA, corresponding to a digestion time of 3–6 min at room temperature. Coverage of the amino acid sequence ranged from 92% for cytochrome c to 71% for BSA, with some missed cleavages observed. Melittin was consumed within 5 s. In contrast, a similar extent of digestion of melittin in a cuvet took 10–15 min. The kinetic limitations associated with the rapid digestion of low picomole levels of substrate were minimized using an integrated digestion bed with hydrodynamic flow to provide an increased ratio of trypsin to sample. This chip design thus provides a convenient platform for automated sample processing in proteomics applications. Copyright © 2000 John Wiley & Sons, Ltd.Keywords
This publication has 34 references indexed in Scilit:
- Microfluidic Devices Connected to Fused-Silica Capillaries with Minimal Dead VolumeAnalytical Chemistry, 1999
- Microfabricated Devices for Capillary Electrophoresis−Electrospray Mass SpectrometryAnalytical Chemistry, 1999
- Integration of Microfabricated Devices to Capillary Electrophoresis−Electrospray Mass Spectrometry Using a Low Dead Volume Connection: Application to Rapid Analyses of Proteolytic DigestsAnalytical Chemistry, 1999
- Microfabricated device coupled with an electrospray ionization quadrupole time-of-flight mass spectrometer: protein identifications based on enhanced-resolution mass spectrometry and tandem mass spectrometry dataRapid Communications in Mass Spectrometry, 1998
- An Integrated Microfluidics-Tandem Mass Spectrometry System for Automated Protein AnalysisAnalytical Chemistry, 1998
- Nanoflow Solvent Gradient Delivery from a Microfabricated Device for Protein Identifications by Electrospray Ionization Mass SpectrometryAnalytical Chemistry, 1998
- Integrated multichannel microchip electrospray ionization mass spectrometry: analysis of peptides from on-chip tryptic digestion of melittinRapid Communications in Mass Spectrometry, 1997
- A Microfabricated Device for Rapid Protein Identification by Microelectrospray Ion Trap Mass SpectrometryAnalytical Chemistry, 1997
- Generating Electrospray from Microchip Devices Using Electroosmotic PumpingAnalytical Chemistry, 1997
- Multichannel Microchip Electrospray Mass SpectrometryAnalytical Chemistry, 1997