Abstract
The primary-parallel-secondary-series LCC (PPSS-LCC) resonant high voltage (HV) power supply with single-phase AC input has been widely applied for medical X-ray application. It requires extremely low ripple (<0.5%) for hundreds of kilovolts output voltage, which is quite challenging. Large capacitance is often utilized at the DC bus to absorb the second-order ripple power introduced by the single-phase input, which deteriorates the power density of the system. To address this issue, this paper proposes an output voltage ripple suppression strategy based on plug-in repetitive control (RPC) combined with pulse-frequency modulation (PFM). It can realize voltage ripple suppression and ZVS of the LCC converter at the same time. The DC-bus capacitance can, therefore, be remarkably reduced. This strategy has the advantages of no additional hardware costs, directly implemented in digital controller, and insensitive to accuracy of system parameters. Experiments are carried out on a practical 120kV/900W HV power supply and confirm the effectiveness of the proposed strategy. The peak-peak output voltage ripple is suppressed to only 0.36%, while the DC capacitance is significantly reduced to 1/12. Comparison of X-ray does prove that the image quality of X-ray machine is consequently greatly improved.
Funding Information
  • National Natural Science Foundation of China (52277062)

This publication has 27 references indexed in Scilit: