Abstract
Hip fractures caused by sideways falls impose a major health and economic burden on older adults, yet hip protectors are often rejected because they are thick, heavy, and uncomfortable. This study used finite element (FE) modeling and drop-tower testing to develop a lightweight two-zone natural rubber hip protector (TZNRHP), comprising a stiff solid-rubber inner zone over the greater trochanter and a closed-cell rubber-foam outer zone. A hyperelastic–viscoelastic FE model was validated against drop-tower tests on a surrogate hip (4.75 kg impactor, 0.52 m drop), and protectors 9–15 mm thick with six ventilation-hole patterns were evaluated (n = 5 repeated impacts; one-way ANOVA with Tukey’s test). The FE model predicted peak forces within 5.4% for all validation conditions. Thickness significantly affected peak force (p < 0.001, η2 = 0.99), and the 11 mm protector was the thinnest design significantly below the 3100 N fracture criterion, reducing the peak force from 6232 ± 160 N (unprotected) to 2785 ± 79 N (55.3% attenuation) while being 28.3% lighter than the original 15 mm design. For a female-specific threshold of 2800 N, the 12 mm design was required. Thirteen 5 mm ventilation holes (8.4% open area) raised the peak force by only 2.6% (2856 ± 73 N; not significantly different from the least-perforated pattern), and the 11 mm protector reduced the maximum femoral von Mises stress from 9.2 to 5.83 MPa. The optimized TZNRHP is a promising lightweight, ventilated hip protector for older adults.
Funding Information
  • Agricultural Research Development Agency (CRP5707011210)

This publication has 46 references indexed in Scilit: