Image artifacts in concurrent transcranial magnetic stimulation (TMS) and fMRI caused by leakage currents: Modeling and compensation
Open Access
- 22 April 2009
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 29 (5) , 1211-1217
Abstract
Purpose: To characterize and eliminate a new type of image artifact in concurrent transcranial magnetic stimulation and functional MRI (TMS‐fMRI) caused by small leakage currents originating from the high‐voltage capacitors in the TMS stimulator system.Materials and Methods: The artifacts in echo‐planar images (EPI) caused by leakage currents were characterized and quantified in numerical simulations and phantom studies with different phantom‐coil geometries. A relay‐diode combination was devised and inserted in the TMS circuit that shorts the leakage current. Its effectiveness for artifact reduction was assessed in a phantom scan resembling a realistic TMS‐fMRI experiment.Results: The leakage‐current‐induced signal changes exhibited a multipolar spatial pattern and the maxima exceeded 1% at realistic coil‐cortex distances. The relay‐diode combination effectively reduced the artifact to a negligible level.Conclusion: The leakage‐current artifacts potentially obscure effects of interest or lead to false‐positives. Since the artifact depends on the experimental setup and design (eg, amplitude of the leakage current, coil orientation, paradigm, EPI parameters), we recommend its assessment for each experiment. The relay‐diode combination can eliminate the artifacts if necessary. J. Magn. Reson. Imaging 2009;29:1211–1217.Keywords
This publication has 16 references indexed in Scilit:
- Interhemispheric Effect of Parietal TMS on Somatosensory Response Confirmed Directly with Concurrent TMS–fMRIJournal of Neuroscience, 2008
- Dorsal Premotor Cortex Exerts State-Dependent Causal Influences on Activity in Contralateral Primary Motor and Dorsal Premotor CortexCerebral Cortex, 2007
- An approach to MRI-based dosimetry for transcranial magnetic stimulationNeuroImage, 2007
- Optimal EPI parameters for reduction of susceptibility-induced BOLD sensitivity losses: A whole-brain analysis at 3 T and 1.5 TNeuroImage, 2006
- Scalp position and efficacy of transcranial magnetic stimulationClinical Neurophysiology, 2005
- On the synchronization of transcranial magnetic stimulation and functional echo‐planar imagingJournal of Magnetic Resonance Imaging, 2003
- Image Distortion Correction in fMRI: A Quantitative EvaluationNeuroImage, 2002
- Optimization of 3-D MP-RAGE Sequences for Structural Brain ImagingNeuroImage, 2000
- Artifacts caused by transcranial magnetic stimulation coils and EEG electrodes in T2∗-weighted echo-planar imagingMagnetic Resonance Imaging, 2000
- Trajectory measurement and generalised reconstruction in rectilinear EPINeuroImage, 2000