Stem–cement porosity may explain early loosening of cemented femoral hip components: Experimental–computational in vitro study
Authors:
Open Access
- 5 December 2006
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 25 (3) , 340-350
Abstract
A combination of laboratory experiment and computational simulation was performed to assess the role of interface porosity on stem migration. The early motion of in vitro prepared cemented femoral components was measured during application of cyclic stair climbing loads. Following testing, transverse sections were obtained and the distribution of pores at the stem–cement interface was determined. Finite element models of cemented stem constructs were developed and a scheme was implemented to randomly assign pores to the stem–cement interface. For a series of 14 in vitro prepared components, pore fractions at the stem–cement interface ranged from 23% to 67%. The majority of pores at the stem–cement interface were less than 1 mm in length with a mean length of 1.27 ± 2.7 mm and thickness of 0.12 ± 0.11 mm. For stems with large pore fractions, pores tended to coalesce in longer extended gaps over the stem surface. Finite element and experimental models both revealed strong positive correlations (r2 = 0.55–0.72; p < 0.0001) between stem–cement pore fraction and stem internal rotation, suggesting that the presence and extent of pores could explain the early motion of the stems. There was an increased volume of cement at risk of fatigue failure with increasing stem migration. Pore fractions greater than 30% resulted in large increases in stem internal rotation, suggesting that attempts to maintain surface porosity at or below this level may be desirable to minimize the risk of clinical loosening. © 2006 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop ResKeywords
This publication has 39 references indexed in Scilit:
- The effects of fluid penetration and interfacial porosity on the fixation of cemented femoral componentsThe Journal of Bone and Joint Surgery. British volume, 2005
- A Rough Surface Finish Adversely Affects the Survivorship of a Cemented Femoral StemClinical Orthopaedics and Related Research, 2005
- Cement-implant interface gaps explain the poor results of CMW3 for femoral stem fixationActa Orthopaedica, 2005
- The influence of surface roughness on stem-cement gapsThe Journal of Bone and Joint Surgery. British volume, 2002
- Fatigue crack growth rate does not depend on mantle thickness: an idealized cemented stem construct under torsional loadingJournal of Orthopaedic Research, 2002
- Reliability of detecting prosthesis/cement interface radiolucencies in total hip arthroplastyJournal of Orthopaedic Research, 2002
- Long-term registration has improved the quality of hip replacement: A review of the Swedish THR Register comparing 160,000 casesActa Orthopaedica, 2000
- Radiographic risk signs for loosening after cemented THA:61 loose stems and 23 loose sockets compared with 42 controlsActa Orthopaedica, 1996
- Coulomb frictional interfaces in modeling cemented total hip replacements: A more realistic modelJournal of Biomechanics, 1995
- Extensive porosity at the cement–femoral prosthesis interface: A preliminary studyJournal of Biomedical Materials Research, 1993