Are Results of Fibrinogen Measurements Transferable?

Authors:
  • NENils Egberg
    Nils Egberg
  • AYAnna Yu
    Anna Yu
  • RMRumjana Merzoug-Djilali
    Rumjana Merzoug-Djilali
  • IBInger Blomberg
    Inger Blomberg
  • GJGun Jörneskog
    Gun Jörneskog
  • MBMargareta Blombäck
    Margareta Blombäck

Abstract
Fibrinogen concentration is routinely measured by several methods and the results may influence diagnostic and treatment strategies. It is therefore necessary that results are compatible and transferable between laboratories. In the present study, it is shown that commonly used immuno-nephelometric methods, a commercial variant of the Clauss clotting rate method, and the classical syneresis method, do not differ significantly using patient material, in the interval 2-12 g/l. A research ELISA method that measures intact fibrinogen is not linearly correlated to the syneresis method. The commutability of available calibrators and reference materials (including the WHO 2nd IS) was only 50-80% except for one of the calibrators for which the virtual concentration coincided with that obtained by the syneresis method.