Five‐minute whole‐heart coronary MRA with sub‐millimeter isotropic resolution, 100% respiratory scan efficiency, and 3D‐PROST reconstruction
Open Access
- 29 July 2018
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 81 (1) , 102-115
Abstract
Purpose: To enable whole‐heart 3D coronary magnetic resonance angiography (CMRA) with isotropic sub‐millimeter resolution in a clinically feasible scan time by combining respiratory motion correction with highly accelerated variable density sampling in concert with a novel 3D patch‐based undersampled reconstruction (3D‐PROST).Methods: An undersampled variable density spiral‐like Cartesian trajectory was combined with 2D image‐based navigators to achieve 100% respiratory efficiency and predictable scan time. 3D‐PROST reconstruction integrates structural information from 3D patch neighborhoods through sparse representation, thereby exploiting the redundancy of the 3D anatomy of the coronary arteries in an efficient low‐rank formulation. The proposed framework was evaluated in a static resolution phantom and in 10 healthy subjects with isotropic resolutions of 1.2 mm3 and 0.9 mm3 and undersampling factors of ×5 and ×9. 3D‐PROST was compared against fully sampled (1.2 mm3 only), conventional parallel imaging, and compressed sensing reconstructions.Results: Phantom and in vivo (1.2 mm3) reconstructions were in excellent agreement with the reference fully sampled image. In vivo average acquisition times (min:s) were 7:57 ± 1:18 (×5) and 4:35 ± 0:44 (×9) for 0.9 mm3 resolution. Sub‐millimeter 3D‐PROST resulted in excellent depiction of the left and right coronary arteries including small branch vessels, leading to further improvements in vessel sharpness and visible vessel length in comparison with conventional reconstruction techniques. Image quality rated by 2 experts demonstrated that 3D‐PROST provides good image quality and is robust even at high acceleration factors.Conclusion: The proposed approach enables free‐breathing whole‐heart 3D CMRA with isotropic sub‐millimeter resolution in <5 min and achieves improved coronary artery visualization in a short and predictable scan time.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (EPSRC EP/P001009/, EPSRC EP/P007619, FONDECYT 1161051, FONDECYT 1161055, EP/P001009, EP/P007619)
- National Institute for Health Research
- King's College Hospital NHS Foundation Trust
- Fondo Nacional de Desarrollo Científico y Tecnológico (1161051, 1161055)
This publication has 47 references indexed in Scilit:
- Prognostic Value of Coronary Magnetic Resonance Angiography for Prediction of Cardiac Events in Patients With Suspected Coronary Artery DiseaseJACC, 2012
- The Feasibility of 350 μm Spatial Resolution Coronary Magnetic Resonance Angiography at 3 T in HumansInvestigative Radiology, 2012
- Prospective respiratory motion correction for coronary MR angiography using a 2D image navigatorMagnetic Resonance in Medicine, 2012
- Accelerated contrast‐enhanced whole‐heart coronary MRI using low‐dimensional‐structure self‐learning and thresholdingMagnetic Resonance in Medicine, 2012
- Detection of Intracoronary Thrombus by Magnetic Resonance Imaging in Patients With Acute Myocardial InfarctionCirculation, 2011
- Nonrigid retrospective respiratory motion correction in whole‐heart coronary MRAMagnetic Resonance in Medicine, 2011
- A Prospective Study for Comparison of MR and CT Imaging for Detection of Coronary Artery StenosisJACC: Cardiovascular Imaging, 2011
- Contrast-Enhanced Whole-Heart Coronary Magnetic Resonance Angiography at 3.0-TJACC, 2009
- Advances in sensitivity encoding with arbitrary k‐space trajectoriesMagnetic Resonance in Medicine, 2001
- Comparison of respiratory suppression methods and navigator locations for MR coronary angiography.American Journal of Roentgenology, 1997