Evaluation of a new rebound tonometer for self-measurement of intraocular pressure
- 27 November 2015
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 100 (8) , 1139-1143
Abstract
Background/aims To compare the accuracy of self-obtained, partner-obtained and trainer-obtained measurements using the handheld Icare Home rebound tonometer with Goldmann applanation tonometry (GAT), and to evaluate the acceptability to subjects of Icare Home measurement. Methods 76 subjects were trained to use Icare Home for self-measurement using a standardised protocol. A prespecified checklist was used to assess the ability of a subject to perform self-tonometry. Accuracy of Icare Home self-measurement was compared with GAT using one eye per subject, randomly selected. Bland-Altman difference analysis was used to compare Icare Home and GAT intraocular pressure (IOP) estimates. Acceptability of self-tonometry was evaluated using a questionnaire. Results 56 subjects (74%, 95% CI 64 to 84) were able to correctly perform self-tonometry. Mean bias (95% limits of agreement) was 0.3 mm Hg (−4.6 to 5.2), 1.1 mm Hg (−3.2 to 5.3) and 1.2 mm Hg (−3.9 to 6.3) for self-assessment, partner-assessment and trainer-assessment, respectively, suggesting underestimation of IOP by Icare Home tonometry. Differences between GAT and Icare Home IOP were greater for central corneal thickness below 500 µm and above 600 µm than data points within this range. Acceptability questionnaire responses showed high agreement that the self-pressure device was easy to use (84%), the reading was quick to obtain (88%) and the measurement was comfortable (95%). Conclusions Icare Home tonometry can be used for self-measurement by a majority of trained subjects. IOP measurements obtained using Icare Home tonometry by self-assessment and third party-assessment showed slight underestimation compared with GAT.Keywords
This publication has 22 references indexed in Scilit:
- Potential effects of systematic errors in intraocular pressure measurements on screening for ocular hypertensionEye, 2013
- Home versus ambulatory and office blood pressure in predicting target organ damage in hypertensionJournal Of Hypertension, 2012
- Intraocular Pressure Control and Long-term Visual Field Loss in the Collaborative Initial Glaucoma Treatment StudyOphthalmology, 2011
- Fluctuation of Intraocular Pressure and Glaucoma Progression in the Early Manifest Glaucoma TrialOphthalmology, 2007
- 24-Hour Monitoring of Intraocular Pressure in Glaucoma Management: A Retrospective ReviewJournal of Glaucoma, 2003
- Factors Related to the 4-Year Risk of High Intraocular PressureArchives of Ophthalmology (1950), 2003
- Twenty-four-Hour Intraocular Pressure Pattern Associated with Early Glaucomatous ChangesInvestigative Ophthalmology & Visual Science, 2003
- The Ocular Hypertension Treatment StudyArchives of Ophthalmology (1950), 2002
- The advanced glaucoma intervention study (AGIS): 7. the relationship between control of intraocular pressure and visual field deteriorationAmerican Journal of Ophthalmology, 2000
- Individual Variability in the Diurnal Intraocular Pressure CurveOphthalmology, 1993