Comparison of glare in YAG-damaged intraocular lenses: Injection-molded versus lathe-cut
Authors:
- Patricia E. BathPBPatricia E. Bath
- Yadavinder DangYDYadavinder Dang
- Wayne H. MartinWMWayne H. Martin
- 1 November 1986
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 12 (6) , 662-664
Abstract
A comparative analysis of YAG laser intraocular lens (IOL) damage was undertaken on injection-molded and lathe-cut IOLs. Damage sites were evaluated with polarized light. A consistent positive polarization was observed in the damage sites of lathe-cut IOLs. A consistent negative polarization was observed in the damage sites of injection-molded IOLs. The presence of positive polarization in IOL damage sites may be correlated with increased potential for glare. Results and clinical implications are discussed.Keywords
This publication has 3 references indexed in Scilit:
- Glare and contrast: Indications for cataract surgeryJournal of Cataract & Refractive Surgery, 1986
- Quantitative concepts in avoiding intraocular lens damage from the Nd:YAG laser in posterior capsulotomyJournal of Cataract & Refractive Surgery, 1986
- Glare and AgeArchives of Ophthalmology (1950), 1960