Abstract
A recognized hazard of prolonged endotracheal intubation is progressive airway occlusion resulting from deposition of secretions on the inner surface of the endotracheal tube (ETT). When volume-controlled ventilation is used, progressive ETT occlusion may be detected by monitoring the difference between peak and plateau airway pressures. In pressure-controlled modes, however, inspiratory airway pressures are preset and thus cannot act as a warning indicator. Instead, changes in delivered tidal volumes may aid the diagnosis of ETT occlusion. To determine whether tidal volume monitoring effectively de-tects progressive ETT occlusion, we mathematically modeled the response of a ventilator operating in pressure-controlled mode to increasing airway resistance. To corroborate our model, we then bench-tested the Siemens 300 and Puritan-Bennett 7200 ventilators by using a test lung and a series of ETTs ranging in size from 9.0 to 3.5 mm inner diameter to simulate progressive occlusion. We found that when pressure-controlled mode was used, progressive ETT occlusion did not reduce delivered tidal volumes until occlusion was nearly complete. We conclude that prolonged use of pressure-controlled mode may allow significant ETT obstruction to build up undetected, risking complete ETT occlusion and complicating the perioperative care of patients ventilated with this mode.