The effect of electrode belt size selection upon evaluation of the distribution of ventilation using electrical impedance tomography
Authors:
- 1 November 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-4
Abstract
Electrical impedance tomography (EIT) is a noninvasive diagnostic method enabling imaging of planar distribution of thorax impedance. Since EIT is not a routine tool in clinical practice, the examination methods have not been standardized yet. EIT system PulmoVista 500 (Dräger Medical, Germany) uses an electrode belt with 16 equally distributed electrodes and with several options for its fastening around the thorax. The manufacturer produces five sizes of the belts that partly overlap. There are several groups of patients that can use both a smaller and bigger belt. The aim of this study is to investigate whether there is a significant difference in evaluation of regional ventilation using EIT in case of using two sizes of electrode belt in the same subject and therefore the even distribution of electrodes is not maintained. Thirty healthy subjects were measured in horizontal and vertical position with two belt sizes separately. Significant differences in distribution of ventilation when comparing two belt sizes were found.Keywords
This publication has 9 references indexed in Scilit:
- Regional ventilation distribution determined by electrical impedance tomography: Reproducibility and effects of posture and chest planeRespirology, 2011
- Spontaneous breathing during high‐frequency oscillatory ventilation improves regional lung characteristics in experimental lung injuryActa Anaesthesiologica Scandinavica, 2010
- Protective ventilation using electrical impedance tomographyPhysiological Measurement, 2007
- Positive end‐expiratory pressure optimization using electric impedance tomography in morbidly obese patients during laparoscopic gastric bypass surgeryActa Anaesthesiologica Scandinavica, 2006
- Electrical Impedance Tomography's Correlation to Lung Volume is Not Influenced by Anthropometric ParametersJournal of Clinical Monitoring and Computing, 2006
- Imbalances in Regional Lung VentilationAmerican Journal of Respiratory and Critical Care Medicine, 2004
- Electrical impedance tomography (EIT): a reviewJournal of Medical Engineering & Technology, 2003
- Preliminary static EIT images of the thorax in health and diseasePhysiological Measurement, 2001
- Impedance imaging of lung ventilation: do we need to account for chest expansion?IEEE Transactions on Biomedical Engineering, 1996